Widely used in systems such as RO, pure water, ultrapure water, circulating water, boiler water, as well as in process fields such as electronics, electroplating, chemistry, food, and pharmaceuticals.
Product Details

Application:

Widely used in systems such as RO, pure water, ultrapure water, circulating water, boiler water, as well as in process fields such as electronics, electroplating, chemistry, food, and pharmaceuticals.

characteristic:

96X48 microcomputer designed conductivity/resistivity transmitter

Large LCD display screen with operation status and special symbol display

■ A set of ON/OFF outputs, high/low point selection, hysteresis with arbitrary program settings

■ With program settings to switch ranges and accept input from bipolar and quadrupole electrodes

The combination key enters the setting and calibration mode design to prevent erroneous changes

Specifications:

■ Measurement range

Cond: 0.00μs/cm~199.9ms/cm

Resi: 0.00~19.99 MΩNaN

Temp.: -10~+110.0℃

■ Resolution

Cond: 0.01μs/cm

Resi: 0.01 MΩNaN

Temp.: 0.1℃

■ Instrument accuracy

Cond: 1%±1 Digit

Resi: 1%±1 Digit

Temp.: 0.2℃±1Digit

Temperature compensation: Linear and nonlinear selection at 25 ℃

Automatic temperature compensation Pt1000, NTC30K dual-mode selection

Manual temperature compensation button setting

Electrode constants 0.01, 0.05, 0.1, 0.5, 250px -1 fixed

0.008-19.99 cm-1 adjustable

Working temperature: 0-50 ℃

■ Large LCD display screen with dual indicators of main parameters and temperature

■ Analog output isolated 0/4~20mA output

Programmable Cond/Resi settings

Relay output high/low point selection, hysteresis arbitrary program setting

A set of ON/OFF outputs

240VAC maximum current 0.5A

■ Power input 100-240VAC wide power supply, 50/60Hz

■ Installation method: Panel installation

Instrument size 48mm × 96mm × 110mm (H × W × D)

■ Hole size 44mm × 92mm (H × W)