1Product Introduction
The aerosol monitor is a highly sensitive continuous monitoring instrument for uranium and plutonium aerosols.The use of strip film filter paper and false coincidence method for energy discrimination to measure aerosol concentration has high sensitivity and fast responseIdentifying Natural FactorsThe advantage of strong ability of radon gas particles.Suitable for underground nuclear facilities, nuclear power plants, reprocessing plants, radioactive workplaces, and emergency monitoring of nuclear accidents.
2Technical features
1)The false coincidence method used to distinguish energy effectively identifies interference from natural radon gas particles, making it particularly suitable for application in high radon environments.
2)Using a dual station measuring device that performs sampling and measurement simultaneously,alphaSpectral resolution up to1-2%Can effectively distinguish between uranium and plutonium aerosols.
3)Using a new type of thin film filter paper for sampling, its surface collection characteristics reach80%The above reduces the interference of radon particles to0.1%following.
4)real-time displayradioactivityAerosol energy spectrum, timely detection of pollutant nuclides.
3Technical indicators
1)Monitoring objects: uranium, plutoniumBaerosol
2)Detector:2000mm2PIPSdetector
3)Measurement range:0.01~105 Bq/m3
4)Detection efficiency:uranium(4π)≥28%;plutonium(4π)≥28%
5)Energy Range:alpha:2Mev-10Mev;B:80Kev-2.5Mev
6)Detection lower limit:
Radon concentration10-20 Bq/m3:
Uranium, plutonium0.01-0.02 Bq/m3 (measure time1h),0.06 Bq/m3 (measure time15 min),3 Bq/m3 (measure time2 min), β~2Bq/m3(measure time15 min-1h),
Radon concentration2-3×103Bq/m3:
Uranium, plutonium0.05-0.08 Bq/m3 (measure time1-2h), β 5-10Bq/m3 (measure time1-2h)
7)Response time:10–240minadjustable
8)Sampling filter paper: specialized film filter paper,filtration efficiency> 99%Step by step movement, wide90mm, Long30m
9)Sampling traffic:60-80 L/min
10)Long term stability:300hThe stability of continuous operation is better than±3%
11)Accuracy of reference source response:< 6%
12)Electromagnetic compatibility performance: compliant with national standards151A
13)Alarm methods: sound and light alarm, software interface alarm
14)Communication interface: network port orRS485
15)Environmental conditions: Temperature0℃-+30℃,humidity85%(25℃)
16)Dimensions and Weight: 527*417*326mm(high*wide*deep), 37Kg
4Product Selection Table
Material Number |
Configuration Instructions |
YQ. RD.1.0049 |
Standard version with small cart, no other options available |