E-mail
shkelan@vip.163.com
-
Phone
13381792575
-
Address
No. 492 Xinchuan Road, Pudong New Area, Shanghai/21-1-208 Changjiang Road, Wuxi/209
Shanghai Kelan Instrument Technology Co., Ltd
shkelan@vip.163.com
13381792575
No. 492 Xinchuan Road, Pudong New Area, Shanghai/21-1-208 Changjiang Road, Wuxi/209
G055 series industrial pH meterMain functions and features
1)Can be compatibleORPMeasurement;2)4.3Inch color display screen;3)Capacitive touch button;4)Switching between Chinese and English menus, and operating software text prompts;5)Recorder function, long-term storage of measurement data;6)Two output currents4-20mA;7)Three sets of relay contact outputs;8)Communication interfaceRS485;9)Equipped with temperature automatic compensation and multiple calibration and correction functions for sensors;10)Selection of AC wide voltage and DC power supply for power supply;11)Novel appearance, good sealing performance, multiple installation methods, etc.
G055 series industrial pH meterTwo main technical parameters
1、 Electronic unit:
1.1Measurement range and error
pH:0.0~14.00pHError ± 0.01;(Compatible)ORPMeasurement ±1999mV±5mV).
temperature-20~130℃ error ±0.5℃。
1.2Resolution:pH0.001,temperature0.1.
1.3Protection level:IP67.
1.4Automatic temperature compensation range:0~100℃。
1.5Two way isolated output current and load:4~20mA(Measure the main value and temperature); load<750Oh.
1.6communication interfaceRS485 Modbus RTU.
1.7Relay contacts:3A240VAC,6A28VDCor120VAC(Three groups).
1.8Power supply:85~260VAC,50~60Hzor24VDC.
1.9Quality: Approximately800g.
1.10Overall dimensionsmm:145×145×138Deep, embedded (disk mounted) installation hole size:130×130contain138×138.
There are multiple installation options available: this table is compatible with wall mounted or can be customized as an all-in-one cabinet machine.
1.11Working conditions: Environmental temperature:0~45℃; Relative humidity: not greater than90%;There is no interference except for the Earth's magnetic field.
2、 Sensor electrode:
Sensor electrodes are consumable parts with certain performance differences, and there are multiple models to choose from. Choosing the right one is important for measurement; The main factors considered include: temperature of the measured medium, acidity or corrosiveness of the medium, ordinary water quality or organic matter, solution viscosity, presence of particulate matter in the solution, and special media; We also need to consider the on-site installation requirements and sensor structure.
2.1Basic electrode types:P026WP3type/P891Type used for measuring ordinary media or wastewater0~60℃,P026GP3Type corrosion-resistant, suitable for measuring in sewage or high temperature conditions0-100℃;
P026CP3type/C491Type used for pure water measurement0~80℃, suitable for power plants, etc.
2.2 ORPPtypeORPElectrode is used forORPmeasurementmVValue.
2.3Special solution medium electrode model:P026GTP3Antimony electrodes are corrosion-resistant and resistant to particulate matter;P026TLP3Special electrodes for desulfurization and denitrification are resistant to viscosity and particulate matter;P 026 ggyp 3Type electrodes are used for measuring media such as organic compounds, oxidizing liquids, and hydrogenation liquids.
2.4Digital Sensor Electrode Series: No display instrument required, direct electrode output4-20mA, andRS485The digital signal is directly connected to the display device.
2.5The above electrodes are subjected to water sample pressure:0.5MPa.
2.6There are various installation methods for electrodes, such as flow type, pipeline type, and insertion type, and special installation methods can be customized.