GXH-3010E Portable Infrared CO2 Analyzer

NegotiableUpdate on 01/20
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
This instrument complies with the national metrological verification regulations of JJG635-2011 "Infrared Gas Analyzer for Carbon Monoxide and Carbon Dioxide", and its main technical indicators meet the technical requirements of the national second level instrument. It can obtain a verification certificate from the Chinese Academy of Metrology (verification conclusion: qualified, this instrument meets the second level technical requirements). This instrument has a built-in zero adjustment filter and a six way valve for switching between zero adjustment and measurement, making it easy and flexible to operate. This instrument can customize the measurement range.
Product Details

1、 Product Overview

This instrument complies with the national standards of GB/T18204.2-2014 "Methods for hygiene inspection in public places - Part 2: Chemical pollutants" and GBZ/T300.37-2017 "Determination of toxic substances in workplace air - Part 37: Carbon monoxide and carbon dioxide".

This instrument complies with the national metrological verification regulations of JJG635-2011 "Infrared Gas Analyzer for Carbon Monoxide and Carbon Dioxide", and its main technical indicators meet the technical requirements of the national second level instrument. It can obtain a verification certificate from the Chinese Academy of Metrology (verification conclusion: qualified, this instrument meets the second level technical requirements).

This instrument has a built-in zero adjustment filter and a six way valve for switching between zero adjustment and measurement, making it easy and flexible to operate.

This instrument can customize the measurement range.


2、 Technical parameters

Measurement principle: Non dispersive infrared analysis method/Non dispersive infrared method (NDIR)

Sampling method: built-in pump suction

Measurement range: 0~0.500% or 0~1.000%

Resolution: 0.001%

Repeatability: ≤ 1% full scale

Zero drift: ≤± 2% at full scale/h

Span drift: ≤± 2% full scale/3h

Temperature additional error: (between 10 ℃ and 45 ℃) ≤± 2% at full scale/10 ℃

Carbon monoxide interference: 1250mg/m3CO ≤± 0.3% at full scale

Preheating time: ≤ 10 minutes

Response time: t0~t90 ≤ 15S

Flow range: (0.5-1.0) L/min

Standard configuration: host, sampler, battery, charger, technical documents, portable case

Power supply: AC/DC dual-use, 220AVC (± 10%) or built-in rechargeable battery

Dimensions: 210 × 165 × 85 (mm)

Weight: ≤ 2kg