HGSZ221The fully automatic acid value tester is based onGB/T 28552reachGB/T 264Standard design and development, suitable for acid value analysis and measurement of samples such as transformer oil, turbine oil, and fire-resistant oil. Through the comprehensive application of mechanical, optical, and electronic technologies, automatic functions such as sample weighing, heating reflux, titration, determination of titration endpoint, and printing of measurement results can be achieved. Widely used in fields such as electricity, chemical industry, and environmental protection.
Main technical parameters
Applicable standards:GB/T 28552-2012GB/T 264-1983
Titration accuracy:
The acid value is0.001~0.100mgKOH/gBetween, ≤±0.005mgKOH/g
The acid value is0.100~0.300mgKOH/gBetween, ≤±0.008mgKOH/g
The acid value is0.300~4.000mgKOH/gBetween, ≤±0.014mgKOH/g
Repeatability:
The acid value is0.001~0.100mgKOH/gBetween, ≤0.010mgKOH/g
The acid value is0.100~0.300mgKOH/gBetween, ≤0.016mgKOH/g
The acid value is0.300~4.000mgKOH/gBetween, ≤0.026mgKOH/g
Test scope:0.001~4mgKOH/g
resolving power:0.001mgKOH/g
ambient temperature/Humidity:5℃~40℃/≤85%RH(No condensation)
Power Supply:AC220V±10% 50Hz±5%
Power: ≤500W
External dimensions:264×350×395mm
main features
■ True color GUI touch screen interface, intuitive display and easy operation
The instrument is equipped with a high-precision weighing balance and weighing and testing are carried out on the same testing platform without the need for manual switching
Samples and various chemical reagents can be automatically injected to avoid contact between humans and chemical substances
■ It has a self checking function, which can detect important components during the testing process and facilitate maintenance
The built-in water circulation of the condenser tube does not require an external water source, making the test bench simpler and the operation of the test personnel more convenient
The instrument can calibrate each pump, ensuring the accuracy of instrument testing
The instrument automatically deducts the background values of ethanol and indicator, resulting in more accurate results
Automatic control of the experimental process, automatic prompts, and automatic alarms after the experiment is completed, without the need for manual intervention
■ It has remote communication function, can connect to the network, and experimental data can be uploaded and downloaded (optional)
■ Sample automatic injection and weighing module (optional)