Phone
15588680396
-
Address
No. 23, Siliu North Road, Licang District, Qingdao City
Qingdao Juchuang Environmental Protection Group Co., Ltd
15588680396
No. 23, Siliu North Road, Licang District, Qingdao City
1、 Product Overview
JCY-13S (X) type resistance capacitance method humidity detector is a high-precision portable integrated flue gas humidity and flow rate detector. This instrument uses the humidity sensitive capacitance method to measure the moisture content in flue gas and the S-shaped pitot tube method to measure the flue gas flow rate in pipelines. It can detect various parameters such as flue gas flow rate, flue gas flow rate, standard dry flow rate, dynamic pressure, static pressure, and flue gas temperature of boilers, industrial furnaces, and exhaust ducts. It is widely used in environmental protection, occupational health, labor, safety supervision, military, scientific research, education and other fields.
2、 Execution standards
GB/T 11605-2005 "Methods for measuring temperature and humidity"
JJG 518-1998 "Verification Regulations for Pitot Tubes"
GB/T 16157-1996 "Determination of particulate matter and sampling methods for gaseous pollutants in exhaust from stationary sources"
3、 Technical features
1. Adopting original imported humidity sensors with high measurement accuracy, corrosion resistance, and long service life
2. Adopting imported high-precision micro differential pressure sensors, with small sensor drift and high measurement accuracy
3. Adopting a unique temperature and humidity correction compensation algorithm to eliminate the influence of flue temperature and pressure on measurement results, resulting in higher measurement accuracy
4. Equipped with a rechargeable lithium battery, it can work continuously for more than 4 hours on a single charge
5. The sensor surface is double layered with dust filtration, effectively protecting the measurement results from the influence of dust
6. The sensor surface has a heating function to prevent condensation and effectively protect the measurement results
7. Adopting an integrated design to reduce external interference and make it easy to use
8. Automatically calculate parameters such as average gas flow rate, average pressure, and flue gas flow rate
9. Equipped with automatic zero point correction and software calibration functions to ensure measurement accuracy
10. It has the function of setting up smoke ducts and automatically recommends the number and distance of measurement points
11. Large capacity data storage, capable of storing 100 sets of data files
12. Large size, wide temperature and high brightness display screen display
13. Built in Bluetooth module, optional Bluetooth printer for data printing
4、 Technical parameters
|
TECHNICAL INDEX |
Parameter range |
resolution |
accuracy |
|
humidity |
0~60)% |
0.01% |
Not exceeding ±2% |
|
Smoke dynamic pressure |
0~2500) Pa |
0.01Pa |
Not exceeding ±2% |
|
Smoke static pressure |
(-35~35) kPa |
0.01 kPa |
Not exceeding ±4% |
|
flue gas temperature |
0~400)℃ |
1 ℃ |
Not exceeding ±3 ℃ |
|
atmospheric pressure |
(60~130) kPa |
0.1 kPa |
Not exceeding ±4% |
|
Smoke velocity |
(0~45) m/s |
0.1 m/s |
Not exceeding ±5% |
|
worktemperature |
less than180℃ (Note:180This instrument cannot be used under working conditions above the specified temperature |
||
|
Dimensions (length x width x height) |
1650mm×145mm×85mm |
||
|
response time |
<30S |
||
|
Temperature control temperature |
(0-120)℃ |
||
|
Sampling tube length |
1.3Rice (customizable) |
||
|
Overall Power Consumption |
<60W |
||
|
operation temperature |
(-20-60)℃ |
||
|
working voltage |
built-in batteryDC24V 3AVoltage adapter |
||