Acetic acid, also known as acetic acid or glacial acetic acid, chemical formulaCH3COOHIt is a kind oforganicMonobasic acid, forvinegarMain components. Pure anhydrous acetic acid (glacial acetic acid) is a colorless hygroscopic solid with a freezing point of16.6℃(62℉)After solidification, it becomes a colorless crystal with weak acidity in its aqueous solutioncorrosiveStrong, steam has irritating effects on the eyes and nose.
Borui SiBRS-HD-500Type Online Acetic Acid Concentration-The fully automatic acetic acid detector is an online, high-precision concentration measuring instrument developed based on the principle of critical angle refraction detection. The online detection of concentration is not affected by the color, turbidity, or viscosity of the material; Not affected by bubbles, solid impurities, and crystals in the material; Not affected by changes in liquid pressure, sudden changes in flow rate, or turbulence phenomena; The optical components used in the sensor probe are integrated in design, with direct conversion and processing of photoelectric signals. They are consumable free, maintenance free, and drift free, making them suitable for pipeline installation or installation on the wall of the container, as well as other installation methods. It has the advantages of simple and convenient operation, low maintenance, and reliable work. The concentration detection value can be displayed on site, and the detection data can be obtained through4-20mAAnalog signal orRS485Communication signal connectionPLCandDCSAccurate automatic temperature compensation ensures that the concentration of the solution can be objectively measured in different measurement environments. The fully cast fiberglass shell and fully shielded steel plate inner wall are designed, and the instrument's automatic cleaning system automatically cleans the prism and self diagnostic function promptly eliminates faults, ensuring the accuracy of online measurement. Change the extensive production management method that has relied on manual measurement, manual recording, and manual control for a long time, comprehensively improve the stability of concentration, and eliminate concentration imbalance caused by human control. It has a significant effect on improving product quality and enhancing work efficiency.

BRS-HD-500Type Online Acetic Acid Concentration-Main technical parameters of fully automatic acetic acid detector:
|
measurement range |
concentration 0.0 - 500 |
resolution |
Brix 0.1% |
refractive index 1.33300~1.46500 |
refractive index 0.00001 |
||
temperature 0.0 ~ 100.0℃ |
temperature 0.1℃ |
||
|
measurement accuracy |
Brix±0.1% |
ambient temperature |
0.0 - 60.0℃ |
refractive index±0.00003 |
sample size |
Immersion measurement |
|
temperature±0.5℃ |
pressure resistance |
1MPa |
|
measure time |
default 5 Second measurement 1 time |
Protection level |
Measurement sectionIP67 |
Installation method |
Threads, flanges, chucks, etc. Support customization. |
||
|
output |
modeldraftoutput:24V 4-20mA rightshouldconcentration 0.0 -500(branchHold steadymanufacture) |
||
Digital output:RS485 | |||
Other output methods can be customized. | |||
input |
DC +24V±10% current<100mA |
||
|
Optional parts |
Automatic cleaning device, using compressed air or high-pressure water and ultrasonic waves to automatically clean the prism surface Surface, suitable for dirty liquids, prevents dirt from adhering to the detection surface, ensuring measurement accuracy. |
||
The concentration of a solution refers to a certain amount of solution(Or solvent)The amount of solute contained in it,The traditional method of analyzing and measuring solution concentration is entirely manual, which is easily affected by factors such as sampling environment, temperature changes, and manual operation, and cannot effectively guarantee the stable quality indicators of the product. Resulting in lagging production process control, low automation level, and inability to achieve real-time fully automatic monitoring and control of concentration.