Overview
The online analyzer for trace water in chlorine gas is a trace water analysis instrument developed by our company specifically for detecting the moisture content of highly corrosive gases such as liquid chlorine. $r $n adopts the P2O5 electrolysis method, which complies with national standards such as GBT 5138-2021 Industrial Liquid Chlorine and HGT3941-2007 Determination of Moisture Content in Industrial Liquid Chlorine by Electricity Method. It can measure trace moisture content in various corrosive gases such as chlorine, hydrogen fluoride, and hydrogen bromide (excluding gases that react with phosphoric acid such as ammonia).
Product Details

Electrolytic dew point meter, is a trace water analysis instrument developed by our company specifically for detecting the moisture content of highly corrosive gases such as liquid chlorine.

adoptThe P2O5 electrolysis method complies with national standards such as GBT 5138-2021 Industrial Liquid Chlorine and HGT3941-2007 Determination of Moisture Content in Industrial Liquid Chlorine by Electricity Method,measurableChlorine gas, hydrogen fluoride, hydrogen bromide and other corrosive substancesgas(Excluding gases such as ammonia that react with phosphoric acid)Medium and trace moisture content.

Typical ApplicationsField:

At the bottomChlor alkali industry

At the bottomMeasurement of Trace Water in Various Corrosive Gases

Online analyzer for trace water in chlorine gasInstrument features:

At the bottomSplit type, display unit and detection unit are designed separately to prevent corrosion caused by sample gas

At the bottomThe materials used for components such as the air circuit of the entire machine are corrosion-resistant

At the bottomLarge screen dot matrix formatLCDDisplay, display units are optionalppm

At the bottomAutomatic real-time recording of measurement data

At the bottomGas path and program design, sensor sealing protection, stable overall performance

Following standards:

At the bottomGB/T 5832.1-2016Gas Analysis Determination of Trace Moisture Part 1: Electrolytic Method

At the bottomGBT 5138-2021 Industrial Liquid Chlorine

At the bottomHGT 3941-2007 Determination of moisture content in industrial liquid chlorine - Electricity method

Online analyzer for trace water in chlorine gasPrinciple Description

The characteristic of using materials such as phosphorus pentoxide to absorb moisture and decompose into polar molecules, thereby accumulating charges on the electrode, when the analyzed sample gas enters the electrolytic cell, the moisture in the gas is coated on the surface of the probeP205 hygroscopic agent absorbs and is electrolyzed into hydrogen and oxygen by the direct current voltage applied to the probe electrode, and discharged with the sample gas.The reaction process during measurement is as follows:P2O5+3H2O → 2 H3PO4 Moisture absorption process4H3PO4 → 2P2O5+6H2 ↑+3O2 ↑ Electrolytic process During the electrolysis process, the sample gas generates an electrolysis current. According to Faraday's law of electrolysis and the gas state equation, the electrolysis current generated under certain temperature, pressure, and flow conditions is proportional to the water content in the gas. As long as the magnitude of the electrolysis current is measured, the water content of the gas can be determined.

Technical Specifications

At the bottomAnalysis principle: Principle of electrolytic method for phosphorus pentoxide

At the bottomrangeScope:0-2000ppm(The middle range can be customized)

At the bottomessence degreesoil5%FS

At the bottomheavy again Sex:≤±0.5%F. S

At the bottomstable fixed Sex:±2%F.S/7d

At the bottomdivide distinguish Rate:1ppm

At the bottomResponse time:T90≤10min

At the bottomWorking environment: Temperature -25℃~+55℃humidity≤90%RH(No condensation)

At the bottomSample gas pressure:0.05 MPa ≤ inlet pressure ≤ 0.1 MPa (outlet must be at atmospheric pressure)

At the bottomSample gas flow rate:100ml/min

At the bottomPower supply220VAC±10%,50Hz±5%(Charging power supply); The instrument comes with a rechargeable battery

At the bottomData recording:3000article

At the bottomDisplay unit:Volume ratio(ppm/v)%

At the bottomAlarm method: beep sound alarm, alarm limit can be set arbitrarily within the range