Online total alkalinity analyzer

NegotiableUpdate on 03/29
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
PACON4900 online total alkalinity analyzer is a compact, easy to operate, and precise online analyzer provided by JENSPRIMA, used for automatic online detection of alkalinity and quality control of boiler water. The principle of titration color change method is to measure high/low range by selecting different reagents. Low maintenance and low reagent consumption, capable of long-term continuous operation. Standard 4-20mA and RS485 Modbus outputs can be easily integrated into your existing control system. Measurement parameter: alkalinity (p-value)
Product Details
Online total alkalinity analyzer

Measurement method: titration color change method
Measurement range: TP5050: 0.2-5mmol/L
TP5200:0.8-20mmol/L
Accuracy: ± 5% of the selected reagent value
Repeatability: ± 2.5% of the selected reagent value
Reagent consumption: 0.05-0.6ml/analysis per session
Reagent expiration date: 2 years from factory
Analysis time: about 3 minutes, depending on the alkalinity concentration and the set flushing time
Measurement interval: 5-360min optional
Automatic cleaning: The instrument will automatically clean after each measurement
Calibration: Manual calibration of service menu
Environmental temperature: 10-45 ℃
Water sample temperature: 5-40 ℃
Water sample pressure: 0.5-5bar, recommended 1-2bar
Water quality requirements: clear, colorless, no suspended solids, no bubbles, pH>8.2
Power supply: 100-240VAC, 50/60Hz
Power consumption: 20W (working), 3.5W (standby)
Inlet/outlet: 6mm hose, with open outlet and no back pressure discharge
Current output: 0/4-20mA
Relay output: 4-channel passive NC/NO relay output, 250VAC/1A
External input: 2-channel programmable input contacts
Protection level: IP65
Size: 300 × 300 × 200mm
Weight: Approximately 4Kg



33-4900-00

PACON 4900 online alkalinity analyzer

50-4905-01

TP5050 alkalinity: 0.2~5mmol/L

50-4920-01

TP5200 alkalinity: 0.8~20mmol/L

50-5000-10

Spare parts package, including peristaltic pump head (including pump tube), sealing ring, agitator, reagent bottle connecting tube, recommended to be replaced once a year