Frog Vision New GenerationPPB grade optical dissolved oxygen sensorBased on the principle of fluorescence quenching, blue light is irradiated onto a fluorescent oxygen molecule probe to excite red light. Since oxygen molecules are the quenching agent of the fluorescent probe, they can carry away energy, so the time and intensity of the excited red light are inversely proportional to the concentration of oxygen molecules. The dissolved oxygen sensor calculates the concentration of oxygen molecules by measuring the phase difference between the excited red light and the reference light, and combining it with a temperature compensated 3D calibration data model.
PPB grade optical dissolved oxygen sensoradvantage:
Smart sensor, all calibration parameters are stored inside the sensor, plug and play.
Optical technology eliminates the need for frequent replacement of membranes or replenishment of electrolyte solutions.
No preheating required, fast response speed, with a maximum response time of 30 seconds (T90).
The measurement process does not consume oxygen.
PTFE support material, resistant to certain solvents, with small drift and easy storage.
Our self-developed membrane, oxygen probe, and key components are all controllable.
RS485 digital output, supporting Modbus protocol.
• SS316,PVC, All titanium alloy shell, with multiple options available.
Technical Specifications
Measurement method: Fluorescence quenching method
Measurement range 0-2000ppb (dissolved oxygen)
Resolution of 0.1ppb
Accuracy ± 1 ppb or 3% reading, whichever is greater
Repeatability ± 0.5ppb or 1.5% reading, whichever is greater, for 24 hours
Sample temperature range: 0~45 ℃
Detection limit 0.8ppb
Calibration at 1 point (zero point), in 99.999% nitrogen gas
Use water depth/pressure 1-20bar (absolute pressure)
Power supply 5-12V 25mA (normal)
Output RS485, Modbus RTU protocol
Material: Titanium alloy (optional), SS316 (default), PVC
Size length 141.9mm (excluding cable connectors), outer diameter 27mm
CIP cleaning support
Response time 30 seconds T90 (in gas), 150 seconds T90 (in liquid), oxygen free to aerobic
Service life * Sensor 2 years or more, sensing membrane 2 years or more
Maintenance and calibration frequency * sensor for 6 months
Note: * The lifespan, maintenance, and calibration frequency are for reference only and depend on the actual usage environment.