TK1300 AC/DC current sensor calibration device

NegotiableUpdate on 05/27
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

TK1300 AC/DC current sensor calibration device ◉ TK1300 is a portable instrument designed specifically for detecting the ratio of DC current sensors, the ratio and deviation of AC current sensors. The built-in DC current standard source has an uncertainty level of 0.005, and the uncertainty level of DC secondary signal measurement is also 0.005. It can calibrate DC current sensors with an uncertainty level of 0.05 or below. The built-in AC current standard source has an uncertainty level of 0.03, and the uncertainty level of AC secondary signal measurement is 0.02. It can calibrate AC current sensors with an uncertainty level of 0.2 or below. ◉ Can be widely applied to various industries such as metrology, electricity, rail transit, military industry, instrument manufacturing, scientific research, etc. for the detection standards of AC/DC sensors.

Product Details



Functional Features

The instrument is equipped with a broadband current comparator for current measurement and control, ensuring high stability and accuracy of the current.
Both AC and DC current outputs use the same pair of terminals, avoiding the hassle of switching wires back and forth, making it extremely convenient to use.
◉ Human computer function: Equipped with a large LCD color screen, the measurement value is displayed intuitively, and the mode of touch+number buttons is used to control the measurement value output. The manual inspection meter can be used conveniently.
The AC/DC current source and the secondary signal measurement module of the sensor are isolated from each other, ensuring high system reliability.
◉ Equipped with a dedicated adapter box, it can be connected to various types of secondary signal output ports and converted into dedicated ports to transmit signals to the instrument for measurement.

Direct current I1output

QQ截图20191026162318



AC current I1output

QQ截图20191026162342



DC voltage U2And current I2Measurement/AC voltage U2And current I2measure

QQ截图20191026162432



Auxiliary power supply

Output range: DC ± (5 V~24 V) adjustable, fineness adjustment: 0.1%, maximum load capacity: 1 A
Adjustment method: The output voltage value can be set through buttons, knobs, touch screens, and upper computer software
Protection functions: short circuit protection, overload protection, overheating protection
Measurement function: Voltage, current, power measurement (sensor power consumption measurement)
Voltage/current measurement uncertainty 0.2%, power measurement uncertainty 0.5%

Technical specifications for angle difference measurement

Voltage range: ≥ 20% * RG, current range: ≥ 30% * RG
40 Hz ≤ F ≤ 100 Hz: 300 μrad
100 Hz < F ≤ 400 Hz: 600 μrad
400Hz < F ≤ 1kHz: 1200 μrad