E-mail
guyu.sh@163.com
- Phone
-
Address
Room 701, Building 36, Lane 1600, Ping'an Road, Minhang District, Shanghai
Shanghai Guyu Environmental Protection Technology Co., Ltd
guyu.sh@163.com
Room 701, Building 36, Lane 1600, Ping'an Road, Minhang District, Shanghai

Instrument characteristics
TECHNICAL INDEXTethys400 (UV) Ammonia Nitrogen Online Monitoring Instrument Parameters:
Measurement principle:
The measurement principle of NH3-N is based on the ultraviolet (UV) absorption spectrum of ammonia gas in equilibrium with the liquid phase. Add a small amount of sodium hydroxide to the tested sample to convert NH4+into NH3. Then apply Fourier transform to the spectrum to extract the characteristic ammonia (NH3) absorption signal, so as to measure the NH4+content and obtain the ammonia nitrogen content.
characteristic:
This method has strong selectivity and there is no interference when measuring river water or wastewater;
When measured in the gas phase, the turbidity and chromaticity of the water being measured have no effect;
Wastewater containing suspended particles like activated sludge does not require filtration;
This method was introduced in 1956, but Tethys products used microprocessors to enhance the processing of digital signals;
This method does not require standard solutions;
→ Built in automatic cleaning system, low-cost 5% sulfuric acid cleaning solution is automatically injected into the sample circulation pool for cleaning, while performing automatic zeroing work;
The detection system is in an independent box to maintain good detection results.
Technical parameters:
|
Range of measurement |
Low range 0-100mg/l NH3-N, High range 0-1000mg/l NH3-N |
|
repeatability |
±0.05 mg/l NH3-N |
|
accuracy |
±3--±5%FS |
|
display |
320 × 240 pixel color LCD with backlight - Color Touch Screen Display Interface (CTSDI) |
|
Power Supply |
90~264V AC/50~60Hz,12V~15V DC,3A |
|
ambient temperature |
0℃-50℃ |
|
signal output |
4-20mA isolated output, 12 bits, 500 Ω Max |
|
Communication output |
RS232- No special software required, compared to Excel®Or MODBUS protocol compatible RS485 MODBUS protocol |
|
CE standard |
Electromagnetic compatibility EN50081-2, EN50082-2, EN55011 |
|
protection grade |
IP65/Nema 4X, galvanized steel |
|
Volume |
520×390×250mm(H×W×D) |
|
Weight |
14kg |