Total Copper and Copper Ion Water Quality Online Analyzer

NegotiableUpdate on 02/27
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
1、 Product Overview: Copper is an essential trace element for the human body, with an estimated daily requirement of 20mg for adults
Product Details
1、 Product Overview
  
Copper is an essential trace element for the human body, with an estimated daily requirement of 20mg for adults. When copper in water reaches 0.01mg/L, it has a significant inhibitory effect on water self purification. Copper is highly toxic to aquatic organisms. Some people believe that the initial toxicity concentration of copper to fish is 0.002mg/L, but it is generally considered safe for fish to have a copper content of 0.01mg/L in water. The toxicity of copper to aquatic organisms is related to its form in water, and the toxicity of free copper ions is much greater than that of complexed copper. The main sources of copper pollution are wastewater discharged by enterprises in electroplating, metallurgy, hardware, petrochemicals, and chemical industries.
  
This analytical instrument is capable of long-term unmanned automatic continuous operation according to on-site settings, and is widely applicable to industrial pollution source discharge wastewater, municipal sewage treatment plant wastewater, environmental quality surface water and other occasions. According to the complexity of the on-site testing conditions, corresponding preprocessing systems can be selected to ensure the reliability of the testing process, accurate testing results, and fully meet the on-site requirements of different occasions.
  
2、 Main features
  
*The unique design enables this product to have lower failure rates, lower maintenance requirements, lower reagent consumption, and higher cost-effectiveness compared to similar products.
*The intelligent fault self diagnosis function makes instrument management and maintenance very convenient.
*Suitable for measuring pollution sources and surface water, as well as for detection in sewage treatment processes.
*It has network functionality and can achieve remote control of data through the network.
*Humanized design enables this product to have lower failure rates, lower maintenance requirements, longer service life, and higher cost-effectiveness and accuracy compared to similar products;
  
3、 Technical specifications

serial number

Specification Name

Technical specifications and parameters

1

test method

Colorimetric measurement

2

measurement range

0~5mg/L(≤0.5 mg/L;>0.5 mg/L)

3

detection limit

0.02

4

resolution

0.001

5

accuracy

≤±10%( ≤0.5mg/L); ≤±5%(>0.5mg/L)

6

precision

≤10%( ≤0.5mg/L); ≤5%(>0.5mg/L)

linearity

≤10%( ≤0.5mg/L); ≤5%(>0.5mg/L)

measurement period

The minimum testing period is 20 minutes, and the digestion time can be set from 5 to 120 minutes based on the actual water sample

9

sampling period

Time interval (adjustable), hourly or trigger measurement mode, adjustable

10

calibration cycle

Automatic calibration (adjustable from 1 to 99 days), manual calibration can be set according to actual water samples

11

maintenance cycle

The maintenance interval is greater than one month, about 30 minutes each time

12

Human machine operation

Touch screen display and command input

13

Self checking protection

Self diagnosis of instrument working status, data will not be lost after abnormalities or power outages; After abnormal reset or power outage, the instrument automatically eliminates residual reactants and resumes operation automatically

14

data storage

Data storage for no less than six months

15

input interface

discrete signal

16

output interface

2-channel RS232 digital output, 1-channel 4-20mA analog output (expandable 2-channel)

17

working environment

For indoor work, it is recommended to maintain a temperature of 5-28 ℃ and a humidity of ≤ 90% (without condensation)

18

Power supply and power consumption

AC230±10% V,50-60Hz,5A

19

size

Height 1500 x Width 550 x Depth 450 (mm)