WB-8501C/WB-8501A axis displacement monitoring instrument

NegotiableUpdate on 03/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
In parameter setting mode, the WB-8501C/WB-8501A axis displacement monitoring instrument displays parameter flags in the channel I display box and corresponding parameter values in the channel II display box.
Product Details

WB-8501C/WB-8501A axis displacement monitoring instrument

Number setting and operation

This instrument has been debugged before leaving the factory, and users can use it directly. If you need to view the factory parameters of the instrument or make modifications, you can perform the following operations.

After power on, the instrument is in operation mode and the panel displays the displacement of each channel.

All parameters of this instrument are set by the "Parameter Settings,+, -, ←" buttons on the panel. Each parameter is represented by corresponding symbols. The order and steps for setting are as follows:

Parameter flag

Parameter Description

Set range

E1–

Channel I alarm I

0~999μm

P1–

Channel I alarm II

0~999μm

L1–

Channel I range

0~999μm

E2–

Channel II alarm I

0~999μm

P2–

Channel II alarm II

0~999μm

L2–

Channel II range

0~999μm

EE-

alarm delay

0~60s

Note: In parameter setting mode, the parameter flag is displayed in the channel I display box, and the corresponding parameter value is displayed in the channel II display box.

Press the "SET" button for about 5 seconds to enter the parameter setting mode. The display box of channel I on the panel shows "E1-", and the display box of channel II shows the factory channel I alarm value "063". Press the "Confirm" button to enter the parameter editing mode. At this time, the low digit will flash. Use the "+" and "-" buttons to add or subtract 1 from this digit, with a range of 0-9; Press the "←" key, the previous digit will flash, and you can use the "+" key to modify this digit; Similarly, select the digit to be modified by pressing the "←" key, then use the "+" and "-" keys to modify this digit, and press the "SET" key to complete the parameter editing. Press the "+" key to continue setting the next parameter. Or press and hold the "SET" button for about 5 seconds to return to normal working mode.

Terminal board wiring diagram and sensor installation

Debugging instructions

The instrument has been calibrated and qualified before leaving the factory, and users can use it directly.

Installation method

The monitoring device is installed in the control room. When installing, insert the monitoring device from the front of the control room, then install the left and right mounting brackets from the back, and tighten the screws. Connect the wires to the back terminal board of the monitoring device as required.

WB-8501C/WB-8501A axis displacement monitoring instrument

OD9200-B/S9210/S9220/S9260/S9860/02/S9820/02/S9881/01/JX5151/JX5121 变送器
VB-Z330/VB-Z310/VB-Z320/MMS3110/MMS3120/MMS3210/MMS3311/MMS3410/VRS20/ 变送器
MLV/D VIB-15a/HR-2000/TR4101/SJ6011/SJ6021/SJ6031 transmitters
TK3-2/TM0520-K/-9292/JX-3B/S21330/S21320/OD9002/V503/V505/CIJ-J30/CIJ-J20/HV-520/VB-Z8901/ DX-02/VRS5000/SE-11/SE-41/PJD/VRS2000 Vibration verification table eddy current probe calibrator
TM0540/3300XL/WIN-400/HY-XY1/OD-9003 (A/B)/S21310/VRS4000/XQ-15/DZ-30 Eddy Current Probe Static Calibration Instrument