Welcome Customer !

Membership

Help

Nanjing Zhuoma Electromechanical Co., Ltd
Custom manufacturer

Main Products:

cep-online>Products

Nanjing Zhuoma Electromechanical Co., Ltd

  • E-mail

    312534289@qq.com 467090947@qq.com

  • Phone

    18761882890/13814089468

  • Address

    15-20 Xingda Road, Nanjing Yangtze River Industrial Development Zone

Contact Now

COD CR online automatic monitoring instrument

NegotiableUpdate on 04/14
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
CODcr online automatic monitoring instrument Chemical Oxygen Demand (COD) is a chemical method used to measure the amount of reducible substances that need to be oxidized in water samples. Under certain conditions, the amount of oxidant consumed to oxidize reducing substances in 1 liter of water sample is used as an indicator to convert it into milligrams of oxygen required per liter of water sample after complete oxidation, expressed in mg/L. It reflects the degree of pollution caused by reducing substances in water. This indicator is also one of the comprehensive indicators of the relative content of organic matter. Ø Performance characteristics: The instrument integrates digestion and colorimetry, with a simple and reliable structure
Product Details

>COD CR online automatic monitoring instrument

Chemical Oxygen Demand (COD) is a chemical method used to measure the amount of reducible substances that need to be oxidized in a water sample. Under certain conditions, the amount of oxidant consumed to oxidize reducing substances in 1 liter of water sample is used as an indicator to convert it into milligrams of oxygen required per liter of water sample after complete oxidation, expressed in mg/L. It reflects the degree of pollution caused by reducing substances in water. This indicator is also one of the comprehensive indicators of the relative content of organic matter.

Ø Performance characteristics

·The digestion colorimetric process of the instrument is completed in one, with a simple and reliable structure, reducing the possibility of errors.

·Adopting imported modified polytetrafluoroethylene transparent hose with a diameter greater than 1.5mm, reducing the probability of particle blockage in water samples.

·It can automatically clean the sampling pipeline to prevent the formation of algae or biofilm.

·After abnormal reset and power failure, the instrument will automatically discharge residual reactants and restore working status.

·Low failure rate, minimal maintenance, and low reagent consumption.

·The intelligent fault self diagnosis function, alarm prompt, instrument management and maintenance are very convenient.

·Adopting multi-channel valves, it has strong anti-corrosion performance, long service life, and is safe and reliable.

Ø Application Fields

Suitable for wastewater with chemical oxygen demand ranging from 10 to 2000 mg/L and chloride concentration below 2.5 g/L Cl ¯, according to user requirements, it can be applied to wastewater with chloride concentration below 20 g/L Cl ¯.

Online monitoring of industrial pollution source wastewater

Online monitoring of industrial process water usage

Online monitoring of water quality at the inlet and outlet of municipal sewage treatment plants

Ø Technical parameters

Model number

DH310C1

Method basis

National standard GB11914-89 "Water Quality - Determination of Chemical Oxygen Demand - ********* Salt Method"

measurement range

10~2000mg/L COD

Accuracy

Indication error ≤± 10%

repeatability error

≤±10%

measurement period

The small measurement cycle is 40 minutes, and according to the actual water sample, the digestion time can be arbitrarily modified within 5-20 minutes.

sampling period

Time interval (adjustable from 20 to 9999min) and hourly measurement mode

Correction

The instrument can perform sensitivity and zero calibration at regular intervals

calibration cycle

Adjustable at any interval and time from 1 to 99 days

maintenance cycle

Usually once a month, about 30 minutes each time

analog output

2-channel 4-20mA

signal interface

RS232 or RS485

Switching quantity

6 sets of relay control, capacity 220VAC/2A

Comparison of actual water samples

Error ≤± 10%

MTBF

≥ 720h/time

Environmental Requirements

Suggested temperature+5-28 ℃; Humidity ≤ 90% (non condensing)

Size

Height 1500 x Width 550 x Depth 450 (mm)

Power supply

AC220±10%V,50±10%Hz,5A

Other

Abnormal alarms and power outages will not result in data loss; Touch screen display and command input;

After abnormal reset and power failure, the instrument will automatically discharge residual reactants from the device and restore its working state

Workstation installation (recommended) plan diagram

an>