HB2000 online ammonia nitrogen analyzer

NegotiableUpdate on 02/10
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The HB2000 online ammonia nitrogen analyzer automatically completes water sample collection through the control of an embedded industrial computer system. The water sample enters the reaction chamber, and after interference is eliminated by a masking agent, the ammonia nitrogen in the form of free ammonia or ammonium ions (NH4+) in the water sample fully reacts with the reaction solution to form a yellow brown complex. The color of the complex is proportional to the content of ammonia nitrogen. The mixed solution after the reaction enters the colorimetric chamber, and the voltage related to absorbance is detected using the photoelectric colorimetric method. After amplification by a signal amplifier, it is transmitted to an embedded industrial computer. After data processing, embedded industrial computers display ammonia nitrogen concentration values and store, process, and transmit data. Suitable for measuring ammonia nitrogen content in domestic sewage, industrial wastewater, and surface water.
Product Details

HB2000 online ammonia nitrogen analyzer

The HB2000 online ammonia nitrogen analyzer automatically completes water sample collection through the control of an embedded industrial computer system. The water sample enters the reaction chamber, and after interference is eliminated by a masking agent, the ammonia nitrogen in the form of free ammonia or ammonium ions (NH4+) in the water sample fully reacts with the reaction solution to form a yellow brown complex. The color of the complex is proportional to the content of ammonia nitrogen. The mixed solution after the reaction enters the colorimetric chamber, and the voltage related to absorbance is detected using the photoelectric colorimetric method. After amplification by a signal amplifier, it is transmitted to an embedded industrial computer. After data processing, embedded industrial computers display ammonia nitrogen concentration values and store, process, and transmit data. Suitable for measuring ammonia nitrogen content in domestic sewage, industrial wastewater, and surface water.

main features

    ◇ Large screen LCD display

    Adopting a large screen LCD display, it can intuitively display the measurement process, measurement results, reports, and measurement result curves.

    ◇ Large capacity electronic hard drive

    Using a large capacity electronic hard drive, it can store 10 years of historical data (calculated based on 24 data records per day).

    ◇ Automatic calibration of working curve

    Regular automatic calibration of the working curve can ensure long-term accurate measurement.

    ◇ Power outage protection

    It has a power-off protection function. If the instrument accidentally loses power during operation, it can automatically restore its original operating state after being powered on again, and the instrument parameters will not be lost after power failure.

    ◇ 4mA~20mA analog interface output

    Compatible with various standard data collectors.

    Provide DB25 printing interface

    Can be connected to an external micro printer to print monitoring data.

    ◇ RS232 communication interface, optional data collector

    The instrument has RS232 communication interface, optional data collector and other accessories, which can accurately transmit measurement results, and also realize remote parameter modification and remote startup.

    ◇ GSM short message remote monitoring

    After selecting the corresponding data collector and GSM communication module accessories, the instrument can be remotely monitored through GSM short message.

technical specifications

  • measuring range
  • Indication error
  • Repeatability error
  • zero drift
  • stability
  • Minimum measurement cycle
  • output signal
  • communication interface
  • maximum power
  • ambient temperature
  • relative humidity
  • atmospheric pressure
  • Power Supply
  • Sampling outlet
  • (0~100) mg/L (segmented optional)
  • ≤2.0mg/L:±0.2mg/L; >2.0mg/L:±10%
  • ≤3%
  • ±3%FS/24h
  • ≤10%/4h
  • 20min
  • 4mA~20mA
  • RS232
  • 300VA
  • 5℃~35℃
  • ≤85%
  • 86kPa~106kPa
  • AC 220V± 10%,50Hz
  • Distance L ≤ 15m, drop H ≤ 6m