-
E-mail
wuym@winsafety.com
-
Phone
18520228916
-
Address
Room 301, Huangpu Trade Plaza, No. 680 Guangxin Road, Huangpu District, Guangzhou City, Guangdong Province
Guangzhou Mingle Instrument Co., Ltd
wuym@winsafety.com
18520228916
Room 301, Huangpu Trade Plaza, No. 680 Guangxin Road, Huangpu District, Guangzhou City, Guangdong Province
Heating solution
DETUHigh humidity and low sulfur flue gas analyzerSolution, namelyThe testo flue gas analyzer is equipped with a testo fully heated sampling system, which includes a thermal sampling tube, a heating handle, and a fully heated sampling hose.
Full heating process:The sampling tube, handle, and hose are heated throughout the entire process to prevent condensation caused by smoke during the sampling processSO2 measurement error.
Constant temperature control:Fully heated flue gas sampling hose, length3 m, Equipped with a temperature control box, it can specify any heating temperature within the range of 50~180 ℃, achieving full temperature control without being affected by heat loss caused by environmental temperature.
Accurate measurement:The use of a fully heated sampling system eliminates the presence of smokeThe error caused by SO2 condensation can be accurately measured by using the professional flue gas analyzer testo 350.
national standard:Measurement of desulfurization backend using heating methodSO2, Meet the requirements of national standard HJ/T 397-2007.
Product advantages:
lHeated industrial flue gas sampling probes can ensure high measurement accuracy, even if the flue gas contains high concentrationsSO2/NO2 gas
lMaximum temperature resistance of sampling probe600°C
lPreheating time approximately20 minutes
leasy to operate
When the smoke contains high concentrations ofWhen NO2/SO2 gas requires high measurement accuracy, or when relevant national standards and regulations require a heated sampling system, it is recommended to use this sampling probe kit. The temperature inside the gas path of the heated sampling probe is constant and higher than the condensation temperature of water vapor, which can prevent the condensation of water vapor in the flue gas in the sampling system.
High humidity and low sulfur flue gas analyzerSet, only compatible withThe testo 350 is easy to carry when used in combination.
Temperature resistance of sampling tubeAt 600 ℃, the sampling tube can be extended up to 3m in length by connecting it. The thermocouple (2.2m) can be installed inside a 2m long sampling tube for measuring smoke temperature. When the smoke at the sampling point contains a lot of dust, a pre filter can be installed at the end of the heated sampling tube.
the testo 350Smoke analysis box
Technical data
O₂measurement
measurement range 0 ~ +25 Vol.%
measurement accuracy ± 0.8% full scale (0~+25 vol.%)
resolution 0.01 Vol.% (0 ~ +25 Vol.%)
response timet₉₀ 20 s (t95)
CO (H₂compensation)
measurement range 0 ~ +10000 ppm
measurement accuracy ± 10% measurement value (+2001~+10000 ppm)
± 5% measurement value (+200~+2000 ppm)
±10 ppm (0 ~ +199 ppm)
resolution 1 ppm (0 ~ +10000 ppm)
response timet₉₀ 40 s
H₂ Display only as an indication
CO low (H₂compensation)
measurement range 0 ~ 500 ppm
measurement accuracy ±2 ppm (0 ~ +39.9 ppm)
± 5% measurement value (+40~+500 ppm)
resolution 0.1 ppm (0 ~ +500 ppm)
response timet₉₀ 40 s
H₂ Display only as an indication
NO measurement
measurement range 0 ~ +4000 ppm
measurement accuracy ± 5% measurement value (+100~+1999 ppm)
± 10% measurement value (+2000~+4000 ppm)
±5 ppm (0 ~ +99 ppm)
resolution 1 ppm (0 ~ +4000 ppm)
response timet₉₀ 30 s
NO low measurement
measurement range 0 ~ +300 ppm
measurement accuracy ±2 ppm (0 ~ +39.9 ppm)
± 5% measurement value (+40~+300 ppm)
resolution ±0.1 ppm (0 ~ +300 ppm)
response timet₉₀ 30 s
NO₂measurement
measurement range 0 ~ +500 ppm
measurement accuracy ±5 ppm (0 ~ +99.9 ppm)
± 5% measurement value (+100~+500 ppm)
resolution 0.1 ppm (0 ~ +500 ppm)
response timet₉₀ 40 s
SO₂measurement
measurement range 0 ~ +5000 ppm
measurement accuracy ± 5% measurement value (+100~+2000 ppm)
± 10% measurement value (+2001~+5000 ppm)
±5 ppm (0 ~ +99 ppm)
resolution 1 ppm (0 ~ +5000 ppm)
response timet₉₀ 30 s