Instrument composition
the testo 350portableIndustrial grade flue gas analyzerComposed of two components.
1,Testo 350 handheld device (standalone product): used for performing measurements, operating settings, and controlling the entire analyzer, with a color display screen, displaying measurement data, connecting to a printer, built-in preset menu, interface navigation: intuitive and simple.
2,Testo 350 analysis box: equipped with maintenance compartment and sensor compartment. The maintenance compartment is used for the maintenance and upkeep of instrument batteries, air pumps, etc., while the sensor compartment facilitates users' maintenance, inspection, and self replacement of sensors: plug and play
Product advantages:
1,measurableO2, CO, NO, NO2, SO2, CO2, H2S, CxHy and other flue gas components
2,Enhanced anti-interferenceSO2/SO2low sensor to eliminate interference from CO
3,Built in high-efficiency gas pretreatment module to achieve gas water separation
4,Pre calibrated sensor, plug and play, reducing costs and increasing efficiency
5,Analyze box series connection and synchronous control:16 analysis boxes (16 sampling points) for simultaneous analysis;
6,Can display multiple concentration units synchronously:ppm , *mg/m3, And mg/m3 (reference oxygen content gas emission concentration)
7,Optimized data interface: withCompatibility with LIMS and other customer platforms;
Technical data
O₂measurement
measurement range 0 ~ +25 Vol.%
measurement accuracy ± 0.8% full scale (0~+25 vol.%)
resolution 0.01 Vol.% (0 ~ +25 Vol.%)
response timet₉₀ 20 s (t95)
CO (H₂compensation)
measurement range 0 ~ +10000 ppm
measurement accuracy ± 10% measurement value (+2001~+10000 ppm)
± 5% measurement value (+200~+2000 ppm)
±10 ppm (0 ~ +199 ppm)
resolution 1 ppm (0 ~ +10000 ppm)
response timet₉₀ 40 s
H₂ Display only as an indication
CO low (H₂compensation)
measurement range 0 ~ 500 ppm
measurement accuracy ±2 ppm (0 ~ +39.9 ppm)
± 5% measurement value (+40~+500 ppm)
resolution 0.1 ppm (0 ~ +500 ppm)
response timet₉₀ 40 s
H₂ Display only as an indication
NO measurement
measurement range 0 ~ +4000 ppm
measurement accuracy ± 5% measurement value (+100~+1999 ppm)
± 10% measurement value (+2000~+4000 ppm)
±5 ppm (0 ~ +99 ppm)
resolution 1 ppm (0 ~ +4000 ppm)
response timet₉₀ 30 s
NO low measurement
measurement range 0 ~ +300 ppm
measurement accuracy ±2 ppm (0 ~ +39.9 ppm)
± 5% measurement value (+40~+300 ppm)
resolution ±0.1 ppm (0 ~ +300 ppm)
response timet₉₀ 30 s
NO₂measurement
measurement range 0 ~ +500 ppm
measurement accuracy ±5 ppm (0 ~ +99.9 ppm)
± 5% measurement value (+100~+500 ppm)
resolution 0.1 ppm (0 ~ +500 ppm)
response timet₉₀ 40 s
SO₂measurement
measurement range 0 ~ +5000 ppm
measurement accuracy ± 5% measurement value (+100~+2000 ppm)
± 10% measurement value (+2001~+5000 ppm)
±5 ppm (0 ~ +99 ppm)
resolution 1 ppm (0 ~ +5000 ppm)
response timet₉₀ 30 s
Technical Specifications
size 330 x 128 x 438 mm
Operating temperature -5 ~ +45 °C
shell ABS
Protection level IP40
Battery Type lithium battery
battery life 5 hours
power supply alternating current100V - 240V (50 - 60 Hz)
DC voltage input 11V - 40V
Dew point calculation 0 °Ctd; 99 °Ctd
Maximum dust load 20 g/m ³ smoke and dust volume
Maximum smoke negative pressure -300 mbar
Maximum positive pressure of flue gas 50 mbar
Trigger signal input voltage5-12 V (rising or falling) pulse frequency>1 s. Load: 5 V/max, 5 mA, 12 V/max.40 mA
Maximum humidity load Dew point temperature at the inlet of the analyzer flue gas+70 ℃
Hose length longest16.2 m (connected by 5 extension hoses)
pump flow rate 1 liter/minute, with flow monitoring
Maximum memory 250000 readings
Storage temperature -20 ~ +50 °C
weight 4800 g
Industrial grade flue gas analyzerModular smoke sampling probe, Long335 mm, Temperature resistance of 1000 ℃, including fixed cone, NiCr Ni (TI) thermocouple, 2.2m long anti NO2/SO2 adsorption sampling hose.