Welcome Customer !

Membership

Help

Guangzhou Mingle Instrument Co., Ltd
Custom manufacturer

Main Products:

cep-online>Products

Guangzhou Mingle Instrument Co., Ltd

  • E-mail

    wuym@winsafety.com

  • Phone

    18520228916

  • Address

    Room 301, Huangpu Trade Plaza, No. 680 Guangxin Road, Huangpu District, Guangzhou City, Guangdong Province

Contact Now

Industrial grade flue gas analyzer

NegotiableUpdate on 03/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The testo350 portable industrial grade flue gas analyzer is a new portable flue gas analyzer that integrates multiple principle sensors such as fixed potential electrolysis, non dispersive infrared, catalytic combustion, etc. Up to 10 types of gas sensors can be selected, and 6 component concentrations can be measured simultaneously. It has functions such as efficient flue gas pretreatment, multi-stage smoke and dust filtration, automatic dilution of high concentration flue gas, and automatic protection of sensors. Its high temperature (up to 1800 ℃) flue gas sampling, sensor anti-interference, and ultra-low concentration measurement functions are particularly outstanding.
Product Details

Instrument composition

the testo 350portableIndustrial grade flue gas analyzerComposed of two components.

1,Testo 350 handheld device (standalone product): used for performing measurements, operating settings, and controlling the entire analyzer, with a color display screen, displaying measurement data, connecting to a printer, built-in preset menu, interface navigation: intuitive and simple.

2,Testo 350 analysis box: equipped with maintenance compartment and sensor compartment. The maintenance compartment is used for the maintenance and upkeep of instrument batteries, air pumps, etc., while the sensor compartment facilitates users' maintenance, inspection, and self replacement of sensors: plug and play

Product advantages:

1,measurableO2, CO, NO, NO2, SO2, CO2, H2S, CxHy and other flue gas components

2,Enhanced anti-interferenceSO2/SO2low sensor to eliminate interference from CO

3,Built in high-efficiency gas pretreatment module to achieve gas water separation

4,Pre calibrated sensor, plug and play, reducing costs and increasing efficiency

5,Analyze box series connection and synchronous control:16 analysis boxes (16 sampling points) for simultaneous analysis;

6,Can display multiple concentration units synchronously:ppm , *mg/m3, And mg/m3 (reference oxygen content gas emission concentration)

7,Optimized data interface: withCompatibility with LIMS and other customer platforms;

Technical data

Omeasurement

measurement range 0 ~ +25 Vol.%

measurement accuracy ± 0.8% full scale (0~+25 vol.%)

resolution 0.01 Vol.% (0 ~ +25 Vol.%)

response timet₉₀ 20 s (t95)

CO (Hcompensation)

measurement range 0 ~ +10000 ppm

measurement accuracy ± 10% measurement value (+2001~+10000 ppm)

± 5% measurement value (+200~+2000 ppm)

±10 ppm (0 ~ +199 ppm)

resolution 1 ppm (0 ~ +10000 ppm)

response timet₉₀ 40 s

H Display only as an indication

CO low (Hcompensation)

measurement range 0 ~ 500 ppm

measurement accuracy ±2 ppm (0 ~ +39.9 ppm)

± 5% measurement value (+40~+500 ppm)

resolution 0.1 ppm (0 ~ +500 ppm)

response timet₉₀ 40 s

H Display only as an indication

NO measurement

measurement range 0 ~ +4000 ppm

measurement accuracy ± 5% measurement value (+100~+1999 ppm)

± 10% measurement value (+2000~+4000 ppm)

±5 ppm (0 ~ +99 ppm)

resolution 1 ppm (0 ~ +4000 ppm)

response timet₉₀ 30 s

NO low measurement

measurement range 0 ~ +300 ppm

measurement accuracy ±2 ppm (0 ~ +39.9 ppm)

± 5% measurement value (+40~+300 ppm)

resolution ±0.1 ppm (0 ~ +300 ppm)

response timet₉₀ 30 s

NOmeasurement

measurement range 0 ~ +500 ppm

measurement accuracy ±5 ppm (0 ~ +99.9 ppm)

± 5% measurement value (+100~+500 ppm)

resolution 0.1 ppm (0 ~ +500 ppm)

response timet₉₀ 40 s

SOmeasurement

measurement range 0 ~ +5000 ppm

measurement accuracy ± 5% measurement value (+100~+2000 ppm)

± 10% measurement value (+2001~+5000 ppm)

±5 ppm (0 ~ +99 ppm)

resolution 1 ppm (0 ~ +5000 ppm)

response timet₉₀ 30 s

Technical Specifications

size 330 x 128 x 438 mm

Operating temperature -5 ~ +45 °C

shell ABS

Protection level IP40

Battery Type lithium battery

battery life 5 hours

power supply alternating current100V - 240V (50 - 60 Hz)

DC voltage input 11V - 40V

Dew point calculation 0 °Ctd; 99 °Ctd

Maximum dust load 20 g/m ³ smoke and dust volume

Maximum smoke negative pressure -300 mbar

Maximum positive pressure of flue gas 50 mbar

Trigger signal input voltage5-12 V (rising or falling) pulse frequency>1 s. Load: 5 V/max, 5 mA, 12 V/max.40 mA

Maximum humidity load Dew point temperature at the inlet of the analyzer flue gas+70 ℃

Hose length longest16.2 m (connected by 5 extension hoses)

pump flow rate 1 liter/minute, with flow monitoring

Maximum memory 250000 readings

Storage temperature -20 ~ +50 °C

weight 4800 g

Industrial grade flue gas analyzerModular smoke sampling probe, Long335 mm, Temperature resistance of 1000 ℃, including fixed cone, NiCr Ni (TI) thermocouple, 2.2m long anti NO2/SO2 adsorption sampling hose.