1、 Product Introduction:
MHWGB9500-IThe wireless high-voltage ratio tester breaks through the traditional structure and is carefully designed and manufactured for measuring the current, ratio, and leakage current of high and low voltage primary and secondary circuits. It adopts the latest CT technology and mask digital integration technology, and consists of a dedicated high-voltage detector, low-voltage current clamp, host, high-voltage insulation rod, monitoring software, etc. Its wireless transmission test data can penetrate obstacles in multi story buildings, with a linear transmission distance of up to 30 meters. If insulation rods are not used, they can also be used as high-precision low-voltage clamp ammeters and leakage ammeters, which can accurately measure currents or leakage currents ranging from 0.01mA to 1000A.
2、 Product features:
1. The current clamp is made of special alloy and adopts the latest CT technology and magnetic shielding technology, which is almost unaffected by external magnetic fields, ensuring high-precision, high stability, and high reliability for continuous monitoring throughout the year.
2. The host has a luxurious blue screen LCD display, which is clear at a glance, and also has a large storage space that can store 3000 sets of data.
3. The monitoring software has online real-time monitoring and historical query functions, dynamic display, waveform indication, transformation ratio, frequency indication, maximum, minimum, and average value indication, alarm value setting and alarm indication functions, historical data reading, viewing, saving, analysis, processing, sorting, curve fitting, reporting, printing and other functions.
4. The high-voltage detector is connected to the insulation rod and can be used for measuring the current of high-voltage lines below 60KV, testing the current size and transformation ratio on both sides of transformers, online current monitoring. Its dedicated high-voltage current clamp can easily clamp or withdraw the tested wire by pressing or pulling out the insulation rod, saving time and speed. It is widely used in substations, power plants, industrial and mining enterprises, testing stations, electrical maintenance departments for current detection, power inspection, field electrical operations, etc. The insulating rod is lightweight, with characteristics such as moisture resistance, high temperature resistance, impact resistance, bending resistance, high insulation, and stretchability.
3、 Product Technical Parameters:
merit can |
Online testing and monitoring of current, transformation ratio, and leakage current in high and low voltage primary and secondary circuits; |
electricity source |
DC6VAlkaline dry battery(1.5V AAA×4) |
test method |
Clamp shapedCT |
transmission method |
Wireless transmission, maximum distance for linear transmission is about30m |
display mode |
LCD:128dots×64dotsBlue screen backlight function, suitable for dim places |
LCDsize |
Display Domain:44mm×27mm |
instrument size |
Host: width, height, thickness75mm×170mm×30mm; |
Jaw size |
High voltage tester: φ48mm |
sampling rate |
2time/second |
measurement range |
High voltage detector:0.1mA~1000ALow voltage current clamp:0.01mA~6A |
resolution |
High voltage tester:0.1mALow voltage current clamp:0.01mA |
change 档 |
0.01mA~1000AFully automatic gear shifting |
Accuracy of a single current test |
0.01mA~299mA:±1%±3dgt |
0.30A~49.9A:±2%±5dgt |
|
50.0A~199.9A:±3%±5dgt |
|
200A~600A:±4%±5dgt |
|
601A~1000A:±5%±5dgt |
|
Accuracy of secondary current testing |
0.01mA~10.00A:±2%±3dgt(23℃±3℃,70%RHBelow) |
data storage |
3000Group, pressHOLDKey holds data and automatically numbers storage (data will not be lost during power failure or battery replacement) |
Line voltage |
60KVThe following circuit tests (must be included)5Operation of section insulation rod |
data retention |
pressHOLDKey holds data,HOLDSymbol display, press againHOLDKey Cancel Keep |
Data retrieval |
pressHOLDkey+POWERThe key can enter the data lookup mode |
Overflow display |
Over range overflow function:“OL A”Symbol display |
No signal indication |
When the host does not receive the transmission signal, it prompts "no signal" |
Auto Shutdown |
Start up approximately15After minutes, the instrument panel will automatically shut down to reduce battery consumption (the automatic shutdown function can be canceled) |
battery voltage |
When the battery voltage is lower than4.8VWhen the low battery voltage symbol is displayed, it reminds to replace the battery |
Instrument quality |
2.8Kg(including insulation rod and battery) |
External interference |
No particularly strong electromagnetic field; none433MHzco-channel interference |
Working temperature and humidity |
-25℃~45℃;80%Rhbelow |
Storage temperature and humidity |
-10℃~60℃;70%Rhbelow |
Insulation rod size |
φ32mm,1m/Festival(5Section) |
Low voltage current clamp lead length |
standard2Meter (can be extended according to the distance from the site to achieve)0~50Remote online monitoring by Mi |
insulation strength |
High voltage detector:AC100kV/rms(第5Between the insulation rod and the core of the high-voltage current clamp) |
knot constitute |
Anti drip type II (high pressure detector) |
Random attachment |
Clamp table:1Item; receiver:1Item; Current clamp:1Item; CD:1Item;RS232Communication line:1Item; Instrument box:1Item; Insulated rod(1Rice):5One; Battery (alkaline dry battery)AAA):8a |