1、 Product Introduction:
ZGF-200kV/5mAThe DC high-voltage generator is based on Chinese industry standardsZBF 24003 – 90According to the requirements of the "General Technical Conditions for Portable DC High Voltage Generators", a new generation of portable DC high voltage generators has been redesigned and manufactured. This DC high voltage generator is suitable for conducting DC high voltage tests and leakage current tests on equipment such as zinc oxide lightning arresters, power cables, generators, transformers, and switches in the power sector and enterprise power departments.
2、 Product features:
1TheZGF-HDC high-voltage generators are smaller in size, not only beautiful, reliable, and fully functional compared to similar products, but also lighter in weight, easier to operate, fully functional, and can be used in the field.
2High precision and accurate measurement. The voltage and current meters are both digital displays, with a voltage resolution of0.1kVThe current resolution is1uAThe voltmeter on the control box directly displays the voltage value applied to the load test sample, without the need for an external voltage divider during use, and the wiring is simple. The instrument has high and low voltage ends for measuring leakage current, and the high voltage end is displayed with a circular shielded digital meter, which is not afraid of discharge shock and has good anti-interference performance, suitable for on-site use.
3The convenient overvoltage setting function adopts a digital dial switch, which can intuitively display the set voltage value, making your operation more flexible. The displayed value unit iskV.
4Added high precision75%VDC-1mAThe function of zinc oxide lightning arrester measurement brings great convenience. A gentle press does not require calculation. There are on the control box of this instrument75%The voltage function key, when conducting lightning arrester tests, when the current rises to1000uAPress and hold0.75UDC-- 1mAThe button displays the values on the voltmeter and ammeter75%After completing the data, the voltage regulator potentiometer should be immediately returned to zero position, and the green button should be pressed immediately to cut off the high voltage and turn off the power switch. Conduct other experiments.
3、 Product Technical Parameters:
ZGF-200kV/5mAParameters of DC high-voltage generator
1Power supply:AC220V±10%,50Hz±1%
2Rated output voltage:60kVThe120kVThe200kVThe300kVThe400kVThe600kVThe800kV
3Rated output current:2mAThe3mAThe5mAThe8mAThe10mA
4Accuracy of output current indication:<1level
5Accuracy of output voltage indication:<1level
6Ripple factor: ≤0.5
7、0.75Multiple output voltage indication accuracy:<1Level, with latch
Specifications and Technical Parameters |
40/4 |
60/2 |
60/5 |
120/2 |
120/3 |
120/5 |
200/2 |
200/3 |
200/5 |
300/2 |
300/5 |
Rated voltage(KV) |
40 |
60 |
60 |
120 |
120 |
120 |
200 |
200 |
200 |
300 |
300 |
Rated current(mA) |
4 |
2 |
5 |
2 |
3 |
5 |
2 |
3 |
5 |
2 |
5 |
Rated power(W) |
160 |
120 |
300 |
240 |
360 |
600 |
400 |
600 |
1000 |
600 |
1500 |
Voltage measurement accuracy |
Digital display meter ±(1.0 %Reading ±0.2KV) |
||||||||||
Accuracy of current measurement |
Digital display meter ±(1.0 %Reading ±2) |
||||||||||
Ripple coefficient |
≤0.5% |
||||||||||
Voltage stability |
Random fluctuations, changes in power supply voltage ±10%Time ≤1% |
||||||||||
overload capacity |
The no-load voltage may exceed the rated voltage10%The maximum charging current for ten minutes is1.5Double rated current |
||||||||||
power supply |
Single-phase AC50Hertz220Voltage ±10% |
||||||||||
working method |
intermittent use |
||||||||||
The longest continuous time is30minute | |||||||||||
working environment |
temperature-10~40℃ |
||||||||||
Relative humidity: Room temperature is25Not greater than at ℃85%(No condensation) | |||||||||||
Altitude:1500Below meters | |||||||||||
With capacitor |
The capacitance of the test sample is unlimited |
||||||||||
Load capacity |
available1.5Charging at times the rated current |
||||||||||
structural characteristics |
Epoxy fiberglass electrical insulation double pressure cylinder |
||||||||||
Air insulation, no leakage concerns | |||||||||||
Characteristics of the operation box |
high precision0.75UDC--1mAOne touch button (accuracy ≤1.0%The most suitable test for zinc oxide lightning arrester |
||||||||||
Overvoltage protection adopts dip code setting, which is clear at a glance | |||||||||||
Dual use international standard chassis for standing and sleeping,More convenient on-site | |||||||||||