PALL Emflon PFR hydrophobic sterilization filter cartridge 0.2um
description
The Emflon PFR hydrophobic sterilization filter is widely used for sterilizing and filtering air and gas in manufacturing and biotechnology processes. It is composed of a double-layer Pal @ PTFE membrane with natural hydrophobicity and chemical inertness, specially designed for efficient removal of contaminating bacteria and toxins.
excellent performance
Whether used continuously in high-temperature gases up to 60C, for respirators up to 80C, or for repeated steam sterilization, the Emflon PFR filter ensures high integrity and long service life, while still possessing high flow rate hydrophobic properties. Therefore, it is suitable for small systems and can reduce energy consumption, making it a very economical hydrophobic filter.
Verification of Systems Science
The accuracy of the Emflon PFR hydrophobic sterilization filter is 0.2pm for liquid and 0.003um for gas, which has been validated by the following model microorganisms
Liquid challenge of defective Pseudomonas aeruginosa at a concentration of 107/cm2
Aerosol challenge of defective Pseudomonas aeruginosa
The aerosol challenge of PP7 bacteriophage
Aerosol challenge of gaseous sodium chloride at a flow rate of 100L/min (validation of filtration accuracy of 0.003um in gas)
• Characteristics
Complete verification file
High flow rate, low pressure difference
structurally stable
Steam resistant, long service life
• Advantages
Ensure high sterilization efficiency and high safety guarantee for Z
Can be used for small systems to reduce installation and operating costs
Ensure integrity and reliability
Provide low-cost filtering solutions
• Technical parameters

PALL Emflon PFR hydrophobic sterilization filter cartridge 0.2um
•Order Information
| Item number1 | nominal length | AB type interface code2 |
| AB05PFR2PVH4 | 127 mm (5 in.) | 2 |
| AB1PFR7PVH4 | 254 mm (10 in.) | 7 |
| AB2PFR7PVH4 | 508 mm (20 in.) | 7 |
| AB3PFR7PVH4 | 762 mm (30 in.) | 7 |
| AB4PFR7PVH4 | 1016 mm (40 in.) | 7 |
The P in the item number represents that it has been optimized and evaluated for use in the pharmaceutical industry (including sealed stainless steel reinforcement rings, suitable for in-situ steam sterilization), has passed 100% integrity testing, and comes with an inspection certificate. V represents optimization for respiratory/air/gas applications. The H4 suffix represents silicone O-ring (standard part). For O-ring codes of other available materials, please refer to the appendix or consult
2The standard product for AB05 is the flat end interface with code 2, while the standard parts for other length filter cartridges are code 7 with tail fins. Both come with double 226 O-rings.
System air flow rate vs. pressure difference