Performance parameters of radial support force tester for vascular stentPerformance parameters of radial support force tester for vascular stent
1Instrument Introduction:
The water bath type radial support force tester can test the waterproof pressure grip head for a long time37In a water bath at ℃ (or other desired temperature), it is suitable for constant temperature radial force testing of temperature sensitive metals or other materials, especially for radial force testing of support products such as nickel titanium alloys and shape memory alloys.
2、 Key pointparameter
parameter item |
Specifications |
control system |
PLC+Windowscontrol system |
operating interface |
12Inch embedded industrial computer, switching between Chinese and English |
test software |
Monitoring and parameter settings, historical data, experimental reports, curve monitoring, software upgrades |
printer |
WiFiConnect printer |
force sensor |
0~100NAccuracy: ±0.5%FS(Optional ±)0.1%FS) |
Test sample length |
0~100mm(Other length specifications can be customized) |
speed range |
0~5mm/s(Can be set) |
test method |
Iris type rolling radial loading device |
constant temperature water bath |
RT~50Temperature control accuracy: ±2℃ |
power supply |
AC 220V/50Hz |
Overall dimensions |
680*750*1100mm |
IIIexecuteStandards and Applicable Scenarios
meet the standards
T/CSBM 0018-2021Test method for shape recovery ability of nickel titanium shape memory alloy self expanding vascular stent
YY/T1660-2019Radial load testing method for balloon dilation and self dilation vascular stents
Typical Applications:
Batch quality inspection of production line;
Third party laboratory certification.
4、 Example of Operation Process
1)If the vascular stent has been loaded, place the vascular stent system in37Cscholar2CSoak in a constant temperature water bath for no less than2minRelease the vascular stent from the sheath in a constant temperature water bath;If the bracket is not loaded, this operation is not required.
2)For self expanding variable diameter vascular stents, it is necessary to evaluate the radial support force of different segments.
3)The temperature of the testing device is maintained at37Cscholar2C;The initial aperture is larger than the natural diameter of the test bracket.
5)Place the released vascular stent in the testing hole of the device at least at a constant temperature2min.
5)According to the status of the vascular stent, set the aperture of the testing device as slowly as possible to reduce it to the preloaded diameter of the vascular stent or to reduce the diameter by at least50%Slowly increase to the unloaded state (recommended moving speed is as follows): 0.1mm/s~0.5mm/s).
6)Measure the force value of expansion or compression corresponding to the expected size of the implanted blood vessel diameter during the testing process, and observe whether the vascular stent can restore its initial shape after the external force is removed.
5Configuration List
host1Taiwan;
test software1Set;
Instruction Manual1copy;
Certificate of Conformity1copy;
Warranty Card1copy;
Delivery Receipt1copy;
nameplate1block;
power cord1root;
hex key1set;
Several brochures;