Side flow end expiratory CO2 sensor

NegotiableUpdate on 03/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The Sifang Optoelectronic Side Flow End Breath CO2 Sensor is a CO2 sensor module designed based on the principle of infrared absorption spectroscopy, used to measure the concentration of CO2; Can be applied to devices such as monitors, nutritional metabolism analyzers, and exercise cardiorespiratory function testers. It has the advantages of high precision, good temperature characteristics, low power consumption, fast response time, compact appearance, and flexible integrated installation.
Product Details

End expiratory flowCO2sensorProduct Features

  1. Autonomous NDIR sensor technology

  2. EMI system anti electromagnetic interference

  3. High signal-to-noise ratio and high detection accuracy

  4. Volume based urine integrated installation

  5. 0.47W low power consumption


End expiratory flowCO2sensorTechnical Specifications:

Detection Principle Non dispersive infrared technology (NDIR)
data output CO2 concentration (mmHg), end expiratory CO2, inhaled CO2, respiratory rate
Testing Scope 0 ~ 150 mmHg
resolution 0.1 mmHg
detection accuracy 0 ~ 40 mmHg: ± 2 mmHg
41~70mmHg: True value ± 5%
71~100mm Hg: True value ± 8%
100~150mmHg; True value ± 10%
Respiratory rate range 0~150BPM
Respiratory rate accuracy ±1 breath
stability Short term drift: not exceeding 0.8mmHg (4H)
Long term drift: Maintain accuracy specifications within 120H
input voltage 5.00V (±5%)
average current <100mA
peak current <200mA
Initialization time Display within 15 seconds, full specification accuracy within 2 minutes (25 ℃)
compensation Air pressure: 400 mmHg -850mmHg
Working temperature and humidity Operation: 0C-55 ℃, 10-909% RH (non condensing)
Storage: -40 ℃~70 ℃,<90% RH (non condensing)
data communication TTL
lifespan ≥ 10 years
size 57.5mm*23.7mm*24mm
*Please refer to the specification sheet for specific parameters. If you need more technical information, please contact us