The 2B 465H process ozone detector in the United States is a microprocessor based high concentration ozone analyzer used to detect ozone concentration in process airflow during water treatment, food processing, and research applications.
Product Introduction of 2B 465H Process Ozone Detector in the United States:
Model 465H is a microprocessor based high concentration ozone analyzer used to detect ozone concentration in process airflow during water treatment, food processing, and research applications. Model
465H is designed to provide accurate and stable readings, and can be maintained and calibrated for a long time without maintenance. The Model 465H can be a 19 '' rack mount or a NEMA 4X wall mounted enclosure.
Model
The sensor module within 465H uses a narrow band detector with spectral sensitivity, which eliminates the adverse effects of UV or humidity caused by the UV band passing through the filter. The instrument consists of two detectors: one for ozone detection and the other for compensating for changes in UV lamp intensity. This unique design provides sensors that allow the instrument to operate for long periods of time without calibration or maintenance.
To prevent operating parameters from exceeding the measurement range, the instrument has a built-in diagnostic system that provides alarm signals. The measured values can be intuitively read out when the instrument is in use. These readings, alarms, self diagnosis, and module design can all run for a long time. In addition, any function can be set or monitored through the front dashboard, and can also be remotely set or read through the bidirectional RS-232 port.
Product features of the 2B 465H process ozone detector in the United States:
Standard one-year warranty
UV absorption method
Microcomputer control
Menu driven software
0-5 to 0-25% w/w range
Built in diagnostics
Self check alarm
Quick response time
Very small zero drift
Includes programmable auto zero function
Temperature and pressure compensation (standard)
Optional Ethernet
Technical parameters of 2B 465H process ozone detector in the United States:
Measurement principle: UV absorption method (Bill Lambert's law)
Range: 0-5% w/w to 0-25% w/w
0-100g/Nm3 to 0-400g/Nm3
Measurement unit:% w/w, g/Nm3
Accuracy: ± 1% of full range
Accuracy/repeatability: ± 0.5% of full range
Display resolution: 0.01% w/w, 0.1g/Nm3
Response time:<30 seconds to 95%
Compensation: pressure, temperature (NTP=273.15K, 760mmHg)
Inlet pressure range: 3.0-30.0 pounds per square inch (PSI)
Gas flow rate: 0.2-2.0 LPM
Temperature range: 5-45 ℃
Dimensions (height x width x length):
Rack mounted -5.22 '' (133mm) x 19 '' (483mm) x 18.6 '' (473mm)
NEMA 4X-19.5’’(495mm)x 15.9’’(396mm)x 6.90’’(175mm)
Weight:
Rack mounted -13.3 pounds (6.0 kilograms)
NEMA 4X-15.0 pounds (6.8 kilograms)
Power supply: 110-240v, 50/60Hz, 74W (very large)
Analog output: 0-5V or 4-20mA isolated output
Status output: System normal, invalid reading, UV lamp malfunction
RS-232 (optional RS-485): 57.6Kbaud, DB-9 connector