Phone
13261731718
-
Address
Mingliu Future Building, Majiabao West Road, Fengtai District, Beijing
Beijing Huayi Tongtai Environmental Protection Technology Co., Ltd
13261731718
Mingliu Future Building, Majiabao West Road, Fengtai District, Beijing
The 3M QUEST Soundpro SE-2 noise spectrum analyzer from the United States is a simplified noise analyzer that, when combined with QSP II application software, is easy to operate and convenient to use.
3M QUEST Soundpro SE-2 Noise Spectrum Analyzer Product Introduction:
In order to simplify noise analysis and eliminate tedious analysis, the newly developed SoundPro SE and DL series noise meters and real-time spectrum analyzers combined with QSP II application software can greatly simplify your further noise analysis!
The SoundPro SE/DL series real-time spectrum and noise analyzer includes the following models:
SoundPro SE-2 real-time spectrum and noise analyzer, SoundPro DL-2 real-time spectrum and noise analyzer
SoundPro SE-2-1/1 Real time Spectrum and Noise Analyzer, SoundPro DL-2-1/1 Real time Spectrum and Noise Analyzer
SoundPro SE-2-1/3 real-time spectrum and noise analyzer, SoundPro DL-2-1/3 real-time spectrum and noise analyzer
SoundPro SE-2-1/3-10 real-time spectrum and noise analyzer, SoundPro DL-2-1/3-10 real-time spectrum and noise analyzer
SoundPro SE-1 real-time spectrum and noise analyzer, SoundPro DL-1 real-time spectrum and noise analyzer
SoundPro SE-1-1/1 Real time Spectrum and Noise Analyzer, SoundPro DL-1-1/1 Real time Spectrum and Noise Analyzer
SoundPro SE-1-1/1-AC300 real-time spectrum and noise analyzer, SoundPro DL-1-1/1-AC300 real-time spectrum and noise analyzer
SoundPro SE-1-1/3 real-time spectrum and noise analyzer, SoundPro DL-1-1/3 real-time spectrum and noise analyzer
General applications of 3M QUEST Soundpro SE-2 noise spectrum analyzer in the United States:
Workplace noise assessment
Enforcement of Noise Regulations
Environmental noise assessment
Selection of hearing protection equipment
Administrative and Engineering Control Selection
Assessment of Work Noise Exposure
3M QUEST Soundpro SE-2 Noise Spectrum Analyzer from the United StatesTechnical Specifications
Standards and certifications | |
Acoustics: |
EN/IEC61672 . IEC60651 . IEC 60684 . IEC 61620 . AN SI S1.4-1983 . ANSI S1.11-2004 |
Europe: |
EMC |
General specifications | |
Key: |
14 buttons |
Display format: |
128 x 64 dot matrix |
Background light: |
optical fiber |
language |
English, Spanish, German, French, Italian, Portuguese |
measurement | |
Measurable parameters: |
SPL, MAX, MIN, Peak, Ln, Leq, Lavg, Sel, TWA, Taktm, DOSE, PDOSE, Ldn, CNEL, Exposure |
|
Scope: frequency
Amplitude:
Large peak value: |
Type 1:3Hz-22.4kHz; Type 2: 20 Hz-8kHz; Filter: 1/1 octave band (option) or 1/3 frequency band (option) Bands: 8 selectable, 100dB dynamic range; RTA: 8 selectable options, 80dB dynamic range Exceeding 3dB of full engraving |
parameter | |
Calculation rights: |
A. C. Z (linear) |
Response time: |
Fast and slow pulses |
Octave frequency filter (option): |
Type 1, ANSI S1.11-2004, EN/IEC61260 |
Type 1: |
Center frequency: (1/3 mode) 33 center frequencies ranging from 12.5Hz to 20kHz; (1/1 mode) 11 center frequencies ranging from 16Hz to 16kHz |
Threshold: |
0-140dB |
Exchange rate: |
3, 4, 5, 6dB |
Operation mode: |
Sound level drive operation/pause, clock/date drive power on and operation during programmed settings, external/pause, keyboard start operation/pause during programmed settings |
reference resources: |
|
SPL: |
114dB |
frequency |
1kHz |
directional: |
When using a free field response microphone: zero angle |
calibration |
The calibration history is retained and the confirmation of calibration after testing is recorded in the calibration history |
Calibrator: |
QC-10(114dB 1kHz),QC-20(94dB 250Hz ,114dB 1kHz) |
Record options |
SD card required |
Summary data and time history data records: |
MAX, MIN, Peak, Ln, Leq/Lavg |
Exceeding sound level data recording: |
2 users can choose Ln |
Memory capacity: |
According to the storage capacity of the inserted SD memory card |
Distance: |
1, 10, 15, 30, 60 seconds, 5, 10, 15, 30, 60 minutes |
special function | |
Reply and delete: |
You can choose from 1 second to 20 seconds |
Storage system: |
A detachable secure digital storage card (SD card) is required for storing multiple periods/research settings for data recording options. The file format is PC compatible binary files and QSP software is needed to translate data documents. |
Interface and Connection | |
SD card: |
Used for data recording, storage setting, storage time/research |
Power socket: |
External power supply 9-16VDC |
AC/DC output: |
3.5mm stereo (Tip AC.Ring 1 DC.Ring 2 Gnd) |
10 pin auxiliary connector: |
RS-232, 3 digital outputs, 1 digital input |
USB: |
Suitable for USB 2.0 and Mini B connectors |
Environmental characteristics |
|
temperature |
|
Operation: |
(<± 0.5dB impact) -10 ℃ to+50 ℃ |
Storage: |
-25 ℃ to+70 ℃ |
Humidity: |
TBD |
Vibration&Vibration: |
Magnetic field: 80A/m, 50/60Hz does not affect |
External field: |
Electric 10V/m, 1kHz modulated, 30MHz to 1GHz,<55dBC |
Electronic features |
|
battery |
4 AA alkaline batteries, capable of continuous use for at least 10 hours, according to settings/options, excluding the use of background light |
External DC power input: |
Voltage 9-16VDC |
Microphone: |
Type 2: QE7052; Type 1: BK1936 (standard) |
Instrument input: |
50K ohms, normal input impedance |
Pre amplifier: |
Directly installed on (0.52 "or 13.2mm) microphone, other sizes require an adapter |
Pre amplification device: |
The preamplifier is detachable |
Cable: |
Can drive up to 30 meters of cable, with negligible signal loss |
Input impedance: |
Greater than 1G ohms, with a load of 1pf |
Signal is limited: |
11VAC |
structural characteristics |
|
Installation of tripod: |
There are triangular bracket screws installed on the back of the instrument |
Shell: |
The steel fiber is filled with ABS/Polycarbomate and has internal EMC shielding |
Size: |
3.1 "width x 11.1" height (including preamplifier but without microphone) x 1.6 "thickness |
Weight: |
0.54 kg (including battery) |