Acidity meter/PH meter

NegotiableUpdate on 05/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Acidity meter/desktop model PHS-3C. This instrument is used in feed and feed additive analysis for sample pretreatment in the determination of vitamin B6 and Salmonella. Instrument features: 31/2-digit LED digital display, manual temperature compensation function, one point or two-point calibration, international standard mark, equipped with our E-201-C pH composite electrode. Technical indicators: 1. Instrument level: 0

Product Details

Acidity meter/desktop Model PHS-3C

The purpose of this instrument in the analysis of feed and feed additives is:

In the determination of vitamin B6 andSalmonellaThis instrument is used for sample pretreatment in the determination process

Instrument features: 3 1/2 digit LED digital display, with manual temperature compensation function, can be calibrated with one or two points, with international standard markings, equipped with our factory's E-201-C pH composite electrode
Technical indicators:
1. Instrument level: 0.01 level
2. Measurement range: pH: (0~14.00) pH mV: (0~± 1999) mV (automatic polarity display)
3. Minimum display unit: 0.01 pH, l mV
4. Temperature compensation range: (0~60) ℃
5. Basic error of electronic unit: pH: ± 0.01 pH mV: ± 1 mV ± 1 word
6. Basic error of the instrument: ± 0.02 pH ± 1 word
7. Input current of electronic unit: not greater than 2 × 10-12A
8. Input impedance of electronic unit: not less than l × 1012 Ω
9. Temperature compensator error: ± 0.01 pH
10. Repeatability error of electronic unit: pH: 0.01 pH mV: l mV
11. Instrument repeatability error: not greater than 0.01 pH
12. Electronic unit stability: 0.01 pH ± 1 word/3h
13. Dimensions 1 × b × h, mm: 260 × 190 × 70
14. Weight: 1.5kg
15. Normal usage conditions
a) Environmental temperature: (5-40) ℃;
b) Relative humidity: not exceeding 85%;
c) Power supply: AC (220 ± 22) V, (50 ± l) Hz;