The GNST-ORP 227 online sensor uses platinum electrode as the indicator electrode, Ag/AgCl reference electrode, and water sample to form the primary battery. Measure the oxidation-reduction potential of platinum electrode relative to the saturated Ag/AgCl reference electrode using an electronic millivoltmeter, and then convert the oxidation-reduction potential of the composition relative to the standard hydrogen electrode as the report result.
GNST-ORP 227 online sensorThe analyzer adopts platinum ring glass electrochemical method, corrosion-resistant shell, built-in NTC temperature sensor, and uses RS485 interface and standard Modbus protocol.
Application field
The instrument is widely used in industries such as environmental surface water, municipal sewage, industrial wastewater, industrial process water, swimming pools, etc. and has received high praise from a large number of users.
Product Principle
GNST-ORP 227 online sensorUsing platinum electrode as the indicator electrode, Ag/AgCl as the reference electrode, and water sample to form the primary battery. Measure the oxidation-reduction potential of platinum electrode relative to the saturated Ag/AgCl reference electrode using an electronic millivoltmeter, and then convert the oxidation-reduction potential of the composition relative to the standard hydrogen electrode as the report result.
And = End + Up
Where:
En - oxidation-reduction potential of the tested water sample, mV;
Eind - measured oxidation-reduction potential of water sample, mV;
Eref - measure the potential of saturated Ag/AgCl sensor at temperature, mV;。
Product Features
(1) Industrial grade composite sensor, suitable for harsh working environments
(2) Built in NTC temperature sensor with an accuracy of ± 0.1 ℃
(3) Corrosion resistant casing, waterproof rating IP68, capable of long-term underwater operation
(4) RS485 communication interface, standard Modbus protocol, easy to integrate
Technical parameters
model |
GNST-ORP 227 |
working principle |
Glass platinum electrochemical method |
measurement range |
(-2000~2000)mv; Temperature: (0~50) ℃ |
repetitiveness |
1mV |
Drift (24 hours) |
<10mV |
resolution |
1mV |
response time |
<10s |
Protection level |
IP68 |
communication interface |
RS485, Standard Modbus protocol baud rate 9600 Address bits Factory default 50 |
Dimensions |
(φ33*224)mm, Cable 10 meters (customizable) |
material |
316L stainless steel/PVC/titanium alloy optional |
weight |
0.5kg |
install thread |
Upper end NPT3/4 thread, lower end can be equipped with NPT3/4 external thread |
Working Voltage |
(9~36)V DC |
energy consumption |
0.2 W |
MTBF |
≥ 1440h/time |
Calibration cycle |
3-6 months |
working environment |
(0~50)℃ |