Japanese Fandao Electronic Dissolved Oxygen Meter MC-8W-S/MC-8W-P

NegotiableUpdate on 03/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The instrument installed on the Japanese Fandao Electronic Dissolved Oxygen Meter MC-8W-S/MC-8W-P $r $n can select the DO (Dissolved Oxygen) - current type for continuous monitoring and control according to the measurement environment. $r $n is very suitable for low concentration distribution measurement values, fluorescence type, and is very suitable for drainage and other injection measurement values
Product Details

Japanese Fandao Electronic Dissolved Oxygen Meter MC-8W-S/MC-8W-P

feature

日本饭岛电子溶解氧气计MC-8W-S/MC-8W-P

MC-8W-S
Fast response at low concentrations, excellent zero point repeatability and stability.
One click replacement of oxygen sensor (Wagner), easy to maintain and low-cost
It can be monitored through serial communication via wireless LAN (sold separately).

日本饭岛电子溶解氧气计MC-8W-S/MC-8W-P

MC-8W-P
Even in wastewater treatment facilities where flow rates may fluctuate, the indicated values are easily stable, allowing for safe monitoring
It is unlikely to be affected by interfering substances such as hydrogen sulfide, allowing for stable measurement values.



● Compact body that can be used anywhere (DIN96 size) ●
You can select the upper and lower limit conditions by operating the panel keys

purpose

Distribution type MC-8W-S (electric sensor)

日本饭岛电子溶解氧气计MC-8W-S/MC-8W-P

Manage boiler water supply
Measure deoxygenated water for food/drinking
Monitor the deoxygenated water production equipment
Measurement of deoxygenated water in semiconductors.

Drill type MC-8W-P (fluorescent sensor)

日本饭岛电子溶解氧气计MC-8W-S/MC-8W-P

Please monitor the wastewater treatment facilities (air tanks) (aerobic/anaerobic tanks, nitrification tanks, nitrification tanks)
Monitor aquaculture ponds (in marine areas)
Monitoring of water pollution in rivers, etc.


Japanese Fandao Electronic Dissolved Oxygen Meter MC-8W-S/MC-8W-P

Product specifications

Product Name and Model Develop Controller MC-8W-S Execute Controller MC-8W-P
measurement method Diaphragm type battery Optics (Fluorescence) (JIS K0102:2016)
how to display LED digital display
measurement range ①0.000-0.999mg/l②1.00-20.00mg
/L (up to 19 ° C * 1)
Automatic switching
0.00-20.00mg/l
accuracy Instrument accuracy (only indicating part)
① ± 0.003mg/L (constant temperature)
② ± 0.03mg/l (constant temperature)
linear
0.00-8.00mg/L:±0.1mg/L(恒温)
8.01-20.00mg/l: ± 0.2mg/l (constant temperature)
Response speed (90% response) Within 15 seconds (20 ° C) Within 45 seconds (20 ° C)
analog output DC 4-20mA * 2 (maximum load resistance 400 Ω)
Range selection [0-20mg/l, 0-10mg/l, 0-1mg/l, 0-0.1mg/l]
Continuous communication RS-232C allows external connections
(Cannot be used simultaneously with wireless [option])
- -
controlled output Up/down 2-point non voltage A contact output, with a contact capacity of 250V AC 2A selected using panel keys from 7 control conditions,
The delay timer function can reduce
(can be from 0 to 9 seconds)
Control output setting range 0.00-20.00mg/l
calibration method
Saturation water calibration can be performed by setting a one click calibration correction factor using air
One click calibration
Self diagnostic function During calibration, automatically diagnose sensor lifespan/instability/calibration defects and display messages Automatically diagnose instability/calibration defects during calibration, and
Display a message during measurement, automatically diagnose faults during calibration process and display a message
Correction coefficient function Correct chloride ions, measure water flow velocity, etc. Correct chloride ions, etc.
Time constant function The measured values are averaged over the set time and display (set range of 0.2 to 10.0 seconds) The measured values are averaged over the set time and display (set range of 1.0 to 10.0 seconds)
Clock function When reading measurement values through an external connection, the data will simultaneously output one year, month, time, and time. - -
Working temperature range Execution controller body: -5 to 50 ° C (no condensation)
Swing and probe: from 0 to 35 ° C
Automatic correction
Sensor cover and probe: 0-50 ° C automatic temperature calibration
cable length Standard 5m Standard 10m
power supply Rated voltage AC100-240V 50/60Hz 20VA
Body size and weight 96 (w) x 208 (d) x 96 (h) mm, approximately 1.5 kilograms
Probe size Approximately 28x 125mm in diameter
Distribution shell: Approximately. 42x 77mm (excluding protrusions)
Approximately 50 x 203mm in diameter
(Internal screw: 1-1/4-11-1/2NPT female wire)

*If the temperature is 20 ° C or higher, the upper limit of the measurement range is 20 ° C to 19.7 mg/L, from 35 ° C to 9.9 mg/L.
*When shipping from our company, you can choose between DC0-1V and 1-5V