Probe for use in hydrogen peroxide environment of Rohatyn HC2-SM-HH

NegotiableUpdate on 03/28
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Probe used in hydrogen peroxide environment: sterilized with hydrogen peroxide. Hydrogen peroxide (H2O2) is used for disinfecting cleanrooms, incubators, and other equipment. In this process, the air humidity is saturated through spray, resulting in the formation of micro gel on all object surfaces. Hydrogen peroxide will kill all microorganisms. Then H
Product Details

Probe used in hydrogen peroxide environment: sterilized with hydrogen peroxide. Hydrogen peroxide (H2O2) is used for disinfecting cleanrooms, incubators, and other equipment. In this process, the air humidity is saturated through spray, resulting in the formation of micro gel on all object surfaces. Hydrogen peroxide will kill all microorganisms. Then H 2 O 2 is mechanically recovered or naturally decomposed into harmless H 2 O or O 2 within 48 hours. The humidity level during the application process is crucial for the effectiveness of the process.

Standard humidity sensors are difficult to accurately measure in high concentration H2O2 environments. Hydrogen peroxide will occupy the micropores on the surface of the sensor to affect water vapor measurement. That's why ROTRONIC has developed a special sensor called HYGROMER ® HH-1。 Compared to standard sensors, the lifespan of sensors is greatly extended. Please refer to the sensor data sheet for the exact technical data of the sensor.

Features:

Measure relative humidity, temperature, and dew point

HYGROMER for hydrogen peroxide environment ® HH-1 sensor

High measurement accuracy

Compliant with FDA CFR21 Part 11/GAMP

Technical Specifications:

precision ± 0.8% RH, ± 0.1 ° C, at 10... 30 ° C
probe type Temperature and humidity probes for hydrogen peroxide applications (H)2O2)
Integrated probe measurement range -50…100 °C / 0…100 %rh
Analog output accuracy ±1 mV
power supply 3.3…5 VDC
sensor Hygromer HH-1
temperature sensor PT100 1/3 DIN Class B
Filter type none
Response time T 63 <15>15>
calculated value Dew point/frost point
Anti H2O2 880 ppm / 1200 mg/m3
current consumption 4.5 mA, 3.3 VDC power supply
Audit Tracking/Electronic Records FDA 21 CFR Part 11 and GAMP compatible
Analog output signal (standard) 0…1 V = 0...100 %RH
0…1 V = -40...60 °C
(default setting, freely scalable via HW4 software and AC3001 cable)
communication interface UART
housing material Stainless steel 1.4301
IP protection rating IP65
CE/EMC certification EMC Directive 2007/108/EC
Storage and Transportation Conditions -50…100 °C