Welcome Customer !

Membership

Help

Wuxi Huikete Measurement and Control Technology Co., Ltd
Custom manufacturer

Main Products:

cep-online>Products
Product Categories

Wuxi Huikete Measurement and Control Technology Co., Ltd

  • E-mail

    hkt@huikete.com.cn

  • Phone

    13921186818

  • Address

    Room 1001, Building 6, Xiexin Center, No. 19 Hefeng Road, Xinwu District, Wuxi City

Contact Now

Rozhnik HC2A-SH/HC2A-S3H high-precision probe

NegotiableUpdate on 03/28
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The high-precision probe HC2A-SH/HC2A-S3H meets the high requirements for measurement accuracy, measuring humidity and temperature, and calculating dew and frost points. Application: HVAC, HVAC, food, building equipment, papermaking, textile, pharmaceutical and other industries Characteristics: • Accuracy: at 23 ° C At 5 ° C
Product Details

The high-precision probe HC2A-SH/HC2A-S3H meets the high requirements for measurement accuracy, measuring humidity and temperature, and calculating dew and frost points.

Application:

HVAC, food, building equipment, paper, textile, pharmaceutical and other industries

Features:

Accuracy: ± 0.5% RH at 23 ° C ± 5 ° C, ± 0.1 ° C

Application range: -50... 100 ° C/0... 99% RH

• Digital interface (UART) and range adjustable analog output: 0... 1 V

Standard output range: 0... 1 V=-40... 60 ° C/0... 99% RH

Calibration at 23 ° C, 10, 20, 30, 40, 50, 60, 70, 80, 90% RH, calibration points at 20, 50, 80% RH

Technical Specifications:

Precision, high-precision mode ± 0.5% RH/± 0.1 ° C, at 10... 30 ° C
At 23 ° C and 10, 20, 30, 40, 50, 60, 70, 80, 90% RH
probe type High Accuracy Standard probe
scope of work -50…100 °C / 0…99%rh
Power supply/power consumption Power consumption 3.3…5 VDC
long-term stability <1 %rh=''>1>
sensor Hygromer HT-1
temperature sensor PT100 1/3 DIN Class B
Probe protection Polycarbonate plastic cover, black
filter element polyethylene
Response time T 63 <15>15>
High wind speed 20 m/s, With filter
computational load Dew point or frost point
current consumption 4.5 mA, 3.3 VDC power supply
Repeatability of AIRCHIP 0.02 %rh / 0.01 °C
develop software ROTRONIC HW4
Adjust the probe through software 1 point&multiple points% rh&° C
SOFTWARE UPDATE Through the Internet and HW4 software
User information From HW4 (Device Management)
Data processing through HW4 Interface cable
The device is password protected yes
User defined analog output have
Analog output signal (standard) 0…1 V = 0…99%rh; 0…1 V = -40…60 °C
Digital output signal (standard) UART
Output signal type 2 x 0…1 VDC
Shell material polycarbonate
cable length Passive cables can be up to 5m long, while active cables can be up to 100m long
Reverse voltage protection mechanical connection
Service interface UART
FDA/GAMP compatibility Compatible with FDA 21 CFR Part 11 and GAMP
size Ø15 x 83 mm
weight 10 g