E-mail
hkt@huikete.com.cn
-
Phone
13921186818
-
Address
Room 1001, Building 6, Xiexin Center, No. 19 Hefeng Road, Xinwu District, Wuxi City
Wuxi Huikete Measurement and Control Technology Co., Ltd
hkt@huikete.com.cn
13921186818
Room 1001, Building 6, Xiexin Center, No. 19 Hefeng Road, Xinwu District, Wuxi City
Suitable for heavy industry applications installed through walls, ROTRONIC industrial probes are particularly suitable for high temperature and harsh industrial environments. Measure humidity and temperature, and calculate dew/frost points.
Application:
Production environment, high temperature environment, industrial manufacturing, drying process, climate chamber
Features:
• Scope of work: -100... 200 ° C/0... 99% RH
• Digital interface (UART) and variable range analog output, 0... 1 V
Standard output range: 0... 1 V=-40... 60 ° C/0... 99% RH
Calibrate 10,35,80% RH at 23 ° C
Technical Specifications:
| Accuracy at 23 ± 5 ° C | ±0.8 %rh / ±0.1 °C |
|---|---|
| Precision, standard precision mode | At 23 ° C/10, 35, 80% RH, ± 0.8% RH/± 0.1 ° C |
| probe type | Industrial probe |
| scope of work | -100…200 °C / 0…99%rh |
| measurement range | -100...200 °C (short-term peak load) / 0...99%rh |
| Power supply/power consumption | 3.3 V ± 0.1V / ~4.5 mA |
| long-term stability | <1 rh=''> |
| sensor | Hygromer IN-1 |
| temperature sensor | PT100 1/3 DIN Class B |
| Probe protection | Filter bracket, brass nickel plated |
| filter element | Order separately based on the application |
| Response time T 63 | <15> |
| High wind speed | 3 m/s |
| computational load | Dew point or frost point |
| Repeatability of AIRCHIP | 0.02 %rh / 0.01 °C |
| develop software | ROTRONIC HW4 |
| Adjust the probe through software | 1 point&multiple points% rh&° C |
| SOFTWARE UPDATE | Through the Internet and HW4 software |
| User information | From HW4 (Device Management) |
| Data processing through HW4 | Interface cable |
| The device is password protected | yes |
| User defined analog output | have |
| Analog output signal (standard) | 0…1 V = 0…99%rh; 0…1 V = -40…60 °C |
| Digital output signal (standard) | UART |
| SENSOR FRAME FOR %RH AND °C SENSOR | Yes |
| Output signal type | 2 x 0…1 VDC |
| cable length | Passive cables can be up to 5m long, while active cables can be up to 100m long |
| Probe material | PEEK |
| Reverse voltage protection | mechanical connection |
| Service interface | UART |
| FDA/GAMP compatibility | Compatible with FDA 21 CFR Part 11 and GAMP |
| size | Ø15 x 100 mm / Ø15 x 250 mm / Ø15 x 400 mm / Ø15 x 550 mm / Ø15 x 700 mm |
| weight | 230 g / 260 g / 290 g / 330 g / 350 g / |