Phone
13732662684
-
Address
No. 519 Kunjia Road, Kunshan City, Jiangsu Province
Suzhou Yuanhong Environmental Protection Technology Co., Ltd
13732662684
No. 519 Kunjia Road, Kunshan City, Jiangsu Province

(1) Working principle: The SY-1057 cyanide online water quality analyzer adopts the measurement method of adding tartaric acid and zinc nitrate to water and heating distillation. Simple cyanide and some complex cyanide in the water sample are distilled out in the form of hydrogen cyanide, which is absorbed by sodium hydroxide and reacts with cyanamide T to generate hydrogen chloride. The intermediate product of red color is generated rapidly by the reaction of chromogenic reagent isoniazid pyrazolone, with maximum absorption at a wavelength of 546nm and stable for a certain period of time.
(2) The SY-1057 cyanide online water quality analyzer has the following basic functions:
-Can set, proofread, and display time, including year, month, day, hour, and minute;
-Alarm and display of system abnormalities such as equipment power failure, instrument leakage, and inability to import samples into the reactor. At the same time, stop running until the system is restarted;
-After each measurement, the pre-treatment device, instrument pipelines, valves, and other components are automatically cleaned;
-Automatic protection for power and water outages, as well as automatic recovery for incoming and outgoing power;
-Regular cleaning and sampling functions;
-Touchscreen viewing is more intuitive and easier to operate;
Measurement method: Isonicotinate barbituric acid spectrophotometric method
Test range: 0-1 mg/L (can be expanded according to requirements)
Detection limit: 0.001 mg/L
Resolution: 0.001mg/L
Accuracy: The indication error shall not exceed 10%
Repeatability: Relative standard deviation not exceeding 5%
Zero drift: ≤± 5%
Range drift: ≤± 10%
Reagent consumption: precise measurement, minimal reagent usage, saving costs for users
User maintenance: Maintenance interval>1 month, approximately 1 hour per month; Easy and convenient maintenance
Environmental impact: Water quality will not be damaged under extreme conditions such as high suspended solids, impurities, and floating debris
Communication interface: RS232 digital interface, 4-20mA analog signal output
Dimensions: Height 1400 x Width 560 x Depth 450mm
Weight: 65 kilograms
Power supply: AC (220 ± 10%) V, (50 ± 10%) Hz, 5A
Environmental temperature: 5-40 degrees Celsius
Environmental humidity: ≤ 90% (non condensing)
Atmospheric pressure: 86 kPa to 106 kPa
Sampling outlet: distance L ≤ 15m, drop height H ≤ 6m