1、 Product Introduction
The LYUG-1 ultrasonic level gauge has built-in temperature compensation, power adaptive, and multiple self-developed patented technologies. It has a brand new signal processing technology, which greatly improves the measurement accuracy of the instrument and has a significant suppression function on interference echoes.
Adopting metal aluminum alloy die-casting shell, it has a beautiful appearance and good protective ability. The instrument adopts industrial isolated power supply, and all input and output lines have lightning protection, overvoltage, and overcurrent protection circuits.
Easy installation, maintenance, and calibration. The product is widely used for level measurement in industries such as petroleum, chemical, water treatment, water conservancy, food, and grain. It has the characteristics of safety, cleanliness, high accuracy, long service life, stability and reliability, and easy installation and maintenance.
Conductive liquid media, including highly corrosive liquids such as acids, alkalis, salts, slurries, waste water, and solid-liquid two-phase suspensions with volumetric flow rates, are widely used in industries such as petroleum, chemical, metallurgical, textile, food, pharmaceuticals, papermaking, as well as environmental protection and water conservancy construction.
2、 Product Features
※High precision, with an accuracy of 0.3%, capable of resisting various interference waves.
※Stable and reliable, procurement of imported components, and selection of highly stable and reliable components.
※High waterproof grade, using die cast aluminum alloy waterproof and explosion-proof shell; Protection level IP65. The explosion-proof level of the instrument reaches Ex d (ia) ia Ⅱ BT4 and can monitor the liquid level of all flammable liquids except for hydrogen H2, carbon disulfide CS2, and acetylene C2H2.
※Low failure rate, easy to install, easy to maintain
Ultrasonic level gauge is a non-contact instrument that does not come into direct contact with liquids and has a low failure rate. At the same time, when installing the instrument, there is no need to clean the tank and it does not affect the normal operation of production.
※Multiple means of protection, all input and output lines have lightning protection, overcurrent, and overvoltage protection functions.
3、 Technical parameters
Measurement range: 0-15m (selected according to the actual measurement range)
Blind spot:0.25m~0.6m
Distance measurement accuracy: 0.3% (standard conditions)
Distance measurement resolution:1mm
Pressure: below 4 atmospheres
Instrument display: Equipped with LCD to display liquid level or spatial distance
Analog output: 4-20mA
Digital output: RS485, Modbus protocol (optional)
Power supply voltage: DC24V/AC220V, built-in lightning protection device
Environmental temperature: -20℃~+60℃
Protection level: IP65