Xiangjie Instrument XJ830S Injection Needle Puncture Force Tester $r $n Xiangjie Instrument XJ830S Injection Needle Puncture Force Tester is specialized in testing the puncture force of syringes and injection needles, and can accurately measure whether the sharpness of the syringe injection needle meets the standard requirements.
Product Details

Introduction:

Xiangjie Instrument XJ830S Injection Needle Puncture Force TesterSpecializing in testing the puncture force of syringes and injection needles, it can accurately measure whether the sharpness of the syringe injection needle meets the standard requirements.

Test principle:

Use a puncture force testing device to evaluate the puncture force of the injection needle by measuring the peak force when the needle passes vertically through the simulated skin at a specified speed.

Xiangjie Instrument XJ830S Injection Needle Puncture Force TesterTest method:

1. Place the tested needle and simulated skin at a temperature of 22 ° C ± 2 ° C for at least 24 hours and conduct testing at the same temperature;

2. According to the standard requirements, clamp the simulated skin of appropriate size onto the fixture without any significant tensile or compressive force applied to the simulated skin.

3. Install the test needle on the device, with its axis perpendicular to the surface of the simulated skin, and the needle tip pointing towards the center of the circular puncture area.

4. Set the moving speed to 100mm/min and start the testing device.

5. During the puncture process on the membrane, peak force is measured or force/displacement maps are recorded simultaneously. Note: Do not use membranes that have been punctured in circular puncture areas.

Xiangjie Instrument XJ830S Injection Needle Puncture Force TesterMain technical parameters

1. Specification: XJ830S

2. Accuracy level: Level 1

3. Load: 50N

4. Effective force measurement range: 0.2/100-100/%;

5. Test force resolution, load of 500000 yards: no files are separated inside and outside, and the resolution remains unchanged throughout the entire process.

6. Effective test width: 110mm (can be widened according to customer requirements)

7. Effective experimental space: 700mm

8. Test speed: 1~500mm/min (adjustable arbitrarily)

9. Speed accuracy: within ± 1% of the displayed value;

10. Displacement measurement accuracy: within ± 1% of the displayed value;

11. Deformation measurement accuracy: within ± 1% of the displayed value;

12. Test bench lifting device: fast/slow speed control,

13. Test bench safety device: electronic limit protection

14. Test bench return: Manual can return to the initial position of the test at high speed, and automatic can return automatically after the test is completed;

15. Overload protection: Automatic protection when the load exceeds 10%;

16. Motor: 400W

17. Host size: approximately 450 × 350 × 1550mm

18. Host weight: approximately 90kg

Xiangjie Instrument XJ830S Injection Needle Puncture Force Testerrelevant standards

1、GB15811

2、ISO7864:2016 EN

3、GB18671 -2009

4、GB2024-2016

5、YY0166-2002

6、YY0043-2016

7. YBB00092004-2015 stainless steel injection needle for pre filled syringes