Shenzhen Green En Environmental Protection Technology Co., Ltd
service@greensgrace.com
3rd Floor, Building 9, Jiuxiangling Industrial Zone, Xinwei Community, Xili Street, Nanshan District, Shenzhen
The GR Zn total zinc water quality online automatic monitoring instrument consists of a sampling system, a metering system, a digestion system, a detection system, and a programmable control system. The water sample is collected through an automatic sampling system, and after passing through a high-precision photoelectric sensing measurement system, it is transported to the digestion system. Masking agent, nitric acid, buffer solution, and zinc reagent are then added to generate a blue complex before and after the reaction. The color depth is proportional to the concentration of zinc in the water sample. The automatic controller system detects the absorbance of the solution after the reaction at a specific wavelength to calculate the concentration of zinc in the water sample.
Can achieve the functions of standard sample verification and spiked recovery;
Capable of remote transmission control, bidirectional communication, and other functions;
Total amount of liquid for a single sample<>
Equipped with automatic protection and cleaning functions for power and water outages;
Equipped with measurement over limit alarm function, supporting remote setting of over limit values;
Adopting high-resolution color touch screen, easy to operate and rich in information;
Integrated digestion/colorimetric module patented technology, complete digestion, compact structure, high efficiency;
High precision infrared visible photoelectric non-contact measurement system, with high measurement accuracy and high operational reliability;
By conducting two colorimetric measurements on the same sample within one analysis cycle, the sample concentration can be calculated, effectively compensating for the interference caused by sample chromaticity and turbidity;
Through independently designed sampling pipelines, sampling, quantification, liquid discharge, and each reagent have independent pipelines to avoid measurement errors caused by contamination.
project |
Technical Requirements |
methodological basis |
HJ 535-2009 HJ/T 101-2003 |
measurement range |
0-1 mg/L, Segmentable, expandable, and customizable |
zero drift |
≤±3% |
Span Drift |
≤±5% |
Minimum detection limit |
0.001mg/L |
resolution |
0.001mg/L |
accuracy |
≤±5% |
repetitiveness |
≤3% |
spike recovery rate |
90%-110% |
measurement period |
The minimum measurement cycle is 30 minutes, and the reaction time can be set according to the actual water sample |
Measurement Mode |
(1min~9999min) Any adjustable time interval mode and hourly measurement mode |
calibration cycle |
(1 day to 99 days) Adjustable at any interval and any time |
maintenance cycle |
Usually once a month, about 30 minutes each time |
signal output |
RS-485/RS-232 digital and 4-20mA analog output |
human-computer interaction |
HMI high-resolution color touch screen Chinese interface command input |
external power supply |
220VAC,50Hz,5A |
environmental adaptability |
Temperature (0-40) ℃; Humidity ≤ 90% (non condensing) |
Overall dimensions |
310mm*254mm*524mm |