The GR Ni total nickel water quality online automatic monitoring instrument consists of a sampling system, a metering system, a digestion system, a detection system, and a programmable control system. The preprocessed water sample is pumped into a special reactor and first reacts with an acidic oxidizing reagent to oxidize all forms of nickel in the water sample into divalent nickel ions. Then, the pH of the solution is adjusted and a masking agent is added. Finally, a color developing agent is added to the mixed solution for color development reaction. Within the measurement range, the absorbance of the color developing solution is proportional to the concentration of nickel ions in the water sample, and the nickel content in the water sample is detected and calculated at a specific wavelength.
Can achieve the functions of standard sample verification and spiked recovery;
Capable of remote transmission control, bidirectional communication, and other functions;
Total amount of liquid for a single sample<>
Equipped with automatic protection and cleaning functions for power and water outages;
Equipped with measurement over limit alarm function, supporting remote setting of over limit values;
Adopting high-resolution color touch screen, easy to operate and rich in information;
Integrated digestion/colorimetric module patented technology, complete digestion, compact structure, high efficiency;
High precision infrared visible photoelectric non-contact measurement system, with high measurement accuracy and high operational reliability;
By conducting two colorimetric measurements on the same sample within one analysis cycle, the sample concentration can be calculated, effectively compensating for the interference caused by sample chromaticity and turbidity;
Through independently designed sampling pipelines, sampling, quantification, liquid discharge, and each reagent have independent pipelines to avoid measurement errors caused by contamination.
project |
Technical Requirements |
methodological basis |
GB/T 11910-1989 |
measurement range |
0-1mg/L, Segmentable, expandable, and customizable |
zero drift |
≤±3% |
Span Drift |
≤±5% |
Minimum detection limit |
0.001mg/L |
Resolution |
0.001mg/L |
accuracy |
≤±5% |
repetitiveness |
≤3% |
spike recovery rate |
90%-110% |
measurement period |
The minimum measurement cycle is 25 minutes, and the digestion time can be set according to the actual water sample |
Measurement Mode |
(1min~9999min) Any adjustable time interval mode and hourly measurement mode |
calibration cycle |
(1 day to 99 days) Adjustable at any interval and any time |
maintenance cycle |
Usually once a month, about 30 minutes each time |
signal output |
RS-485/RS-232 digital and 4-20mA analog output |
human-computer interaction |
HMI high-resolution color touch screen Chinese interface command input |
external power supply |
220VAC,50Hz,5A |
environmental adaptability |
Temperature (0-40) ℃, humidity ≤ 90% (non condensing) |
Overall dimensions |
310mm*254mm*524mm |