The GR-CH2O formaldehyde water quality online automatic monitoring device consists of a sampling system, a metering system, a digestion system, a detection system, and a programmable control system. Collect water samples through an automatic sampling system, and after passing through a high-precision photoelectric sensing measurement system. In the presence of excessive ammonium salts, formaldehyde reacts with acetylacetone to form a yellow compound, whose color depth is proportional to the content of formaldehyde. The instrument detects its absorbance at a specific wavelength to convert the concentration of formaldehyde in the water sample.
Can achieve the functions of standard sample verification and spiked recovery;
Capable of remote transmission control, bidirectional communication, and other functions;
Total amount of liquid for a single sample<>
Equipped with automatic protection and cleaning functions for power and water outages;
Equipped with measurement over limit alarm function, supporting remote setting of over limit values;
Adopting high-resolution color touch screen, easy to operate and rich in information;
Integrated digestion/colorimetric module patented technology, complete digestion, compact structure, high efficiency;
High precision infrared visible photoelectric non-contact measurement system, with high measurement accuracy and high operational reliability;
By conducting two colorimetric measurements on the same sample within one analysis cycle, the sample concentration can be calculated, effectively compensating for the interference caused by sample chromaticity and turbidity;
Through independently designed sampling pipelines, sampling, quantification, liquid discharge, and each reagent have independent pipelines to avoid measurement errors caused by contamination.
project |
Technical Requirements |
methodological basis |
HJ 535-2009 HJ/T 101-2003 |
measurement range |
0-2mg/L, Segmentable, expandable, and customizable |
zero drift |
≤±3% |
Span Drift |
≤±5% |
Minimum detection limit |
0.001 mg/L |
resolution |
0.001mg/L |
accuracy |
≤±5% |
repetitiveness |
≤3% |
spike recovery rate |
90%-110% |
measurement period |
The minimum measurement cycle is 30 minutes, and the reaction time can be set according to the actual water sample |
Measurement Mode |
(1min~9999min) Any adjustable time interval mode and hourly measurement mode |
calibration cycle |
(1 day to 99 days) Adjustable at any interval and any time |
maintenance cycle |
Usually once a month, about 30 minutes each time |
signal output |
RS-485 digital and 4-20mA analog output |
human-computer interaction |
HMI high-resolution color touch screen Chinese interface command input |
external power supply |
220VAC,50Hz,5A |
environmental adaptability |
Temperature (0-40) ℃; Humidity ≤ 90% (non condensing) |
Overall dimensions |
310mm*254mm*524mm |