GE Panametrics liquid ultrasonic flowmeter
Product Specifications
Fluid type
Liquid: Acoustic conductive liquid, including most cleaning liquids, as well as liquids containing small amounts of entrained solids or bubbles
Flow measurement
Related transmission time ™ Mode
size
3 inches to 16 inches (80 millimeters to 600 millimeters) standard optional: 72 inches (1800 millimeters) size available upon request
precision
Meter scale with a reading of ± 0.5% 3 to 4 inches (80 to 100 millimeters)
Range: Metric sizes ranging from 7 to 40 feet per second (2.13 to 12.19 meters per second) and 6 inches to 36 inches (150 to 900 millimeters)
Range: 3 to 40 feet per second (0.91 to 12.19 meters per second) ± 0.9% reading 3 to 4 inches (80 millimeters to 100 millimeters) metric size
Range: 3 to 7 feet per second (0.91 to 2.13 meters per second). The final installation must meet the following conditions: the fluid flow profile has been formed (usually at 10 times the diameter upstream and 5 times the diameter downstream of the straight pipe section), and the fluid is a single-phase fluid. Applications that use pipeline layout to generate eddy currents (such as two out of plane elbows) may require additional straight pipe sections or flow control devices. Larger sized instruments can be provided upon request.
Repeatability
Reading ± 0.25% for metric dimensions of 3 to 4 inches (80 to 100 millimeters), range: 7 to 40 feet per second (2.13 to 12.19 meters per second) for metric dimensions of 6 to 36 inches (150 to 900 millimeters), range: 3 to 40 feet per second (0.91 to 12.19 meters per second), reading ± 0.5% for metric dimensions of 3 to 4 inches (80 to 100 millimeters), range: 3 to 7 feet per second (0.91 to 2.13 meters per second)
Scope (bidirectional)
0.1 to 40 feet per second (0.03 to 12.19 m/s)
Variability (overall)
400:1
Sensor systems and materials
Bundle Waveguide Technology ™ (BWT) System probe and bracket made of 316L stainless steel are optional: alternative cushioning materials can be provided upon request
Sensor temperature range
Normal temperature: -310 ° F to 600 ° F (-190 ° C to 315 ° C) liquid
High temperature: -310 ° F to 1112 ° F (-190 ° C to 600 ° C)