The gelatin online detection analyzer uses the principle of refraction to measure the refractive index of a solution. The instrument is designed with integration and is easy to operate. It can monitor the concentration of gelatin in real-time 24 hours a day and has factory linear calibration and automatic temperature compensation functions.
The importance of gelatin concentration detection is reflected in multiple aspects. Firstly, in terms of food safety and quality control, as gelatin is widely used as a thickener and stabilizer, its concentration directly affects the taste, appearance, and stability of food. Therefore, excessive or insufficient concentration may lead to a decrease in quality or even safety risks. Conducting concentration testing can ensure that food meets relevant regulatory standards, effectively prevent illegal merchants from using inferior gelatin or adulteration, protect consumer rights, optimize production processes, improve production efficiency, and ensure product quality stability.
Secondly, in drug production and quality control, given that gelatin is often used in drug formulations such as capsules, the concentration of gelatin solution is one of the key process parameters for controlling the thickness and physical properties of capsule shells, which directly affect the dissolution and release rate of drugs.
In addition, in other industrial applications such as cosmetics, the concentration of gelatin also affects the performance and quality of the product.Therefore, whether in the food, pharmaceutical or other industrial fields, conducting gelatin concentration testing is crucial. It can ensure stable and consistent product quality, comply with regulatory and standard requirements, prevent adulteration and inferior products, optimize production processes, and ultimately protect the rights and interests of consumers and patients.
Product advantages
The gelatin online detection analyzer is not only a measuring device, but also a powerful data platform. It does not require sample pretreatment and consumables, and can provide measurement data within seconds. Thanks to the long-life light source (100000 hours) and the design without mechanical components, the instrument has reliability and stability. The instrument supports communication with the upper computer, providing 4-20mA&RS485 signals, as well as alarm and control switch output. The powerful data modeling function allows you to customize exclusive data models according to specific needs, achieving deeper data analysis and application.
Multiple installation options (chuck, flange pipe adapter, small diameter adapter) ensure flexible installation methods and can be easily integrated into various pipeline systems. The automatic cleaning component further simplifies maintenance work.
The instrument integrates measurement, display, and control functions, with a user-friendly and intuitive interface. The panel displays key data in real-time, such as concentration, temperature, and refractive index. The functions of fault code prompt, one click zero calibration, and factory reset make maintenance more convenient.
Technical Specifications
Product Model |
CYR-G-Gelatin |
measurement items |
Refractive index, temperature, concentration |
measurement range |
0- 40% |
resolution |
0.01% |
measurement accuracy |
± 0.1% |
measure temperature |
-20 to 100 ℃ |
temperature compensation |
0-100°C |
process temperature |
-20°C up to +100°C |
Prism material |
Optical grade sapphire |
Liquid contact material |
316L optional: Hastelloy C-276, 2205, 20 alloy, titanium, tantalum, etc |
input power |
DC24V |
signal output |
DC4 to 20mA&RS485; 2-channel switch signal output |
Protection level |
IP65 |
Process pressure |
≤ 1 MPa (10 Bar), higher pressure levels can be customized |
Dimensions & Weight |
¢ 109 * 138mm/approximately 1.87KG |
Cleaning options |
Optional automatic cleaning component |
Installation Options |
Chuck, flange pipe adapter, small diameter adapter, material: 304, 316L, 2205, PP or steel lined fluorine |
Other functions |
The operation panel can view data such as concentration, temperature, refractive index, etc., and supports functions such as fault code prompt, one click zero calibration, and factory reset |