Hash MS5100 turbidity/residual chlorine/PH multi parameter water quality analyzer

NegotiableUpdate on 04/14
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
-Hash MS5100 turbidity/residual chlorine/PH multi parameter water quality analyzer product features: $r $n $r $n ● MS5100 multi parameter online analyzer adopts an integrated design and occupies a small area. Continuous monitoring includes residual chlorine, chlorine dioxide, turbidity pH、ORP、 Conductivity and temperature are 7 water quality parameters. The MS5100 multi parameter online analyzer adopts a series flow path to meet customers' water saving needs. The MS5100 multi parameter analyzer is equipped with an automatic residual chlorine cleaning mechanism, which has a longer maintenance cycle
Product Details

Hash MS5100 turbidity/residual chlorine/PH multi parameter water quality analyzer

-Product features:

The MS5100 multi parameter online analyzer adopts an integrated design with a small footprint. Continuous monitoring includes residual chlorine, chlorine dioxide, turbidity pH、ORP、 Conductivity and temperature are 7 water quality parameters.

The MS5100 multi parameter online analyzer adopts a series flow path to meet customers' water saving needs

The MS5100 multi parameter analyzer is equipped with a residual chlorine automatic cleaning mechanism, which has a longer maintenance cycle

MS5100 multi parameter water quality online analyzer with electrochemical flow cell modular structure easy to disassemble and clean

The MS5100 multi parameter online analyzer adopts flange sealing for pH, conductivity, and ORP electrodes, which is easy to disassemble and assemble

MS5100 multi parameter online analyzer supports dynamic water-saving mode, effectively reducing water resource consumption

Hash MS5100 turbidity/residual chlorine/PH multi parameter water quality analyzer in stock

Ordering Guide
model configuration
MS5100.10 MS5100 Standard Edition, (including residual chlorine&pH PCBA+turbidity+temperature, excluding residual chlorine and pH electrode)
MS5100.11 MS5100 Standard Edition+Electrochemical Module (Electrochemical PCBA+Flow Cell+Accessories)
MS5100.12 MS5100 Standard Edition+IO Module (IO PCBA+Accessories)
MS5100.13 MS5100 Standard Edition+Electrochemical Module+IO Module (Electrochemical PCBA+Flow Cell+IO PCBA+Accessories)
(1) The electrodes need to be ordered separately. The electrodes are shipped separately and not with the instrument.
(2) The ORP module is optional and only applicable to instruments installed on models MS5100.11 and MS5100.13.
(3) The instruments all include Modbus communication, and 4-20 mA output is an optional module.
technical indicators
measurement method Residual chlorine/chlorine dioxide: Three electrode constant potential method;
Turbidity: 90 degree scattering method, EPA180.1;
PH: glass electrode method;
ORP: Platinum electrode direct measurement method;
Conductivity: Two pole platinum electrode method;
Temperature: Thermistor NTC 30K;
measurement range Residual chlorine: 0~10 mg/L;
Chlorine dioxide: 0-5 mg/L;
Turbidity: 0~100 NTU;
pH:0.00 ~ 14.00;
ORP:-1200 mV ~ +1200 mV;
Conductivity: 20.0 μ S/cm~2000 μ S/cm;
Temperature: 0 ℃~50 ℃
detection limit Residual chlorine/chlorine dioxide: 0.01 mg/L;
Turbidity: 0.0032 NTU;
Conductivity: 6.0 μ S/cm;
accuracy Residual chlorine: ± 1% or ± 0.01 mg/L (whichever is greater);
Chlorine dioxide: ± 1.5% or ± 0.01 mg/L (whichever is greater);
Turbidity: ± 2% or ± 0.01 NTU (whichever is higher);
(Based on Formazine primary standard solution at 25 ℃)
pH:±0.02(2~12 pH);
ORP:±20 mV;
Conductivity: ± 0.8% F.S;
Temperature: ± 0.4 ℃;
repetitiveness Residual chlorine: 1% or 0.01 mg/L (whichever is greater);
Chlorine dioxide: 1.5% or 0.01mg/L (whichever is greater);
Turbidity: 1% or 0.006 NTU (whichever is higher);
(Based on Formazine primary standard solution at 25 ℃)
pH:0.01;
ORP:10 mV;
Conductivity: 0.4% F.S;
Temperature: 0.2 ℃;
display resolution Residual chlorine/chlorine dioxide: 0.001 or 0.01 mg/L optional
Turbidity: 0.001, 0.01, or 0.1 NTU optional;
pH:0.001 、 0.01 or 0.1 is optional;
ORP: 0.01 or 0.1 mV optional;
Conductivity: 0.01, 0.1 μ S/cm/mS/cm optional;
Total dissolved solids (TDS): 0.01, 0.1 μ g/L/mg/L optional;
Temperature: 0.01 ℃;
response time Electrochemical electrode (pH/ORP/conductivity electrode) T90 response time<3 minutes;
System T90 response time (calculated from the moment the water sample enters the instrument)<10 minutes
data log About 104000 records, 12 months (assuming per minute/record)
Operating Temperature 5~40 ℃ (41~104 ℉), relative humidity ≤ 95%, no condensation
Storage temperature -20~60 ℃ (-4~140 ℉), relative humidity ≤ 95%, no condensation
self-diagnosis Equipped with Prognosys self diagnostic system
Measurement Mode Continuous mode, water-saving mode
altitude Up to 2000 meters (6562 feet)
Power requirements 100 - 240 VAC, 50/60 Hz
rated power 30W
analog output 6-way 4-20 mA analog output
digital output RS485 Modbus
display screen 7-inch color resistive touch screen
shell protection IP65 (with the chassis door closed), for indoor use only
size 452 x 681 x 335 mm (17.8×26.8×13.2 in)
Shipping weight 25 kg (51.1 lb), No water supply
Installation method wall-mounted
*If there are any changes, no further notice will be given.