E-mail
18911571363@163.com
-
Phone
18911571363,13366702447
-
Address
1802, Unit 2, Building 9, Area 4, Courtyard 8, Shunsha Road, Changping District, Beijing
Beijing Zhongqing Guotou Environmental Protection Technology Co., Ltd
18911571363@163.com
18911571363,13366702447
1802, Unit 2, Building 9, Area 4, Courtyard 8, Shunsha Road, Changping District, Beijing
Hash ULR CL17sc 0.008mg online total chlorine analyzer
-Instrument features:
The ULR CL17sc online total chlorine analyzer can accurately detect chlorine content in water above 8ppb, monitor and optimize the dechlorination process.
The online total chlorine analyzer uses DPD colorimetric method to directly measure the total chlorine content in water, which is convenient and fast.
The total chlorine analyzer has the function of accumulating chlorine content, which allows customers to understand the real impact of chlorine content through reverse osmosis membranes, thereby more accurately predicting the efficiency and service life of reverse osmosis membranes.
-Measurement method:
The total chlorine analyzer adopts methods such as N, N-yierji, 4-phenylenediamine photometric method (DPD method) for total chlorine determination
-Application industry:
The ULR CL17sc online total chlorine analyzer is widely used in industries such as wastewater, food and beverage production, pharmaceuticals, power, and semiconductors.
| technical indicators | |
|---|---|
| measurement parameter |
total chlorine |
| measurement method | Total chlorine: N, N-yierji-1,4-phenylenediamine photometric method (DPD method) |
| measurement range | 0.008 - 5 mg/L Cl2 |
| accuracy | ± 5% or ± 0.01mg/L (whichever is greater), 0-4 mg/L; ±10%, 4-5mg/L |
| Detection limit | 0.008 mg/L Cl2 , 25℃ |
| measurement period | 2.5min |
| inlet water pressure | 0.3 - 5.2 bar (4.5 - 75 psig), Supply to Y-shaped filter 0.1 - 0.3 bar (1.5 - 5 psig), Supply to analyzer |
| water inlet | Outer diameter ¼ inch polyethylene pipe, quick connector |
| drain outlet | 1/2 inch outer diameter hose |
| air purge | Optional 3/8-inch quick couplings and fittings; When the maximum value is 1.38 bar, 0.003m3/min |
| Sample Requirements | Y-shaped filter with 40mesh or higher mesh sieve |
| flow velocity | The flow rate through the instrument is 60-200 mL/min |
| sample temperature | 5 - 40 °C |
| Working temperature range | 5 - 40 °C |
| Operating Humidity | 0-90% relative humidity, no condensation |
| interfering factor | Oxidants such as bromide, chlorine dioxide, permanganate, and ozone can cause positive errors. |
| reagent consumption | Use 0.5 L buffer solution and 0.5 L indicator solution within 31 days |
| certification | CE符合: EN 61326-1, CISPR 11, EN 50581 ACMA RCM Korean KC certification North America: FCC Supplier Compliance Declaration, IEC/EN 60529, ICES-003 |
| light source | LED, Measured at 510 nm with an optical path length of 1 cm, |
| waterproof rating | IP66 |
| install | wall-mounted |
| Power requirements (voltage) | 12 VDC, Maximum 400 mA (powered by controller) |
| Dimensions (height x width x depth) | 342 mm x 329 mm x 177 mm |
| weight | 4.1 kg |
| Hash ULR CL17sc 0.008mg online total chlorine analyzerOrdering Guide | |
|---|---|
| Ordering Information | |
| 9790200 | ULR CL17 sc total chlorine analyzer, with riser installation kit, does not include reagents and SC controller |
| 9790300 | ULR CL17 sc total chlorine analyzer, with pressure regulator installation kit, does not include reagents and SC controller |
| consumables | |
| 8560400 | CL17sc piping kit (pre assembled) |
| 8573100-CN | CL17sc Color Pool Cleaning Kit |
| accessories | |
| 8568200 | CL17sc Calibration Kit |
| 8573200 | CL17sc Calibration Kit Supplemental Package |
| 8560500 | CL17sc riser installation kit |
| 8565700 | CL17sc pressure regulator installation kit |
| reagent | |
| 9791100S-CN | ULR CL17 sc Total Chlorine Reagent Kit |