Merck Aervent-50 hydrophobic air filter MTGR85000 MTGR85010, 0.2um, 100 pieces/box. Two side barbed connection ports are used to connect silicone tubes with inner diameters ranging from 6.5 to 9.5mm. Diameter 62mm, height 72mm.

Product Details

Merck Aervent-50 hydrophobic air filterMTGR85000 MTGR85010

0.2um, 100 pieces per box. Two side barbed connection ports are used to connect silicone tubes with inner diameters ranging from 6.5 to 9.5mm. Diameter 62mm, height 72mm.

The Aervent-50 0.2um pore size hydrophobic filter consists of a 50mm diameter AerventPTFE membrane sandwiched between two polypropylene shells. This sterile grade filter can be used for important small-scale gas sterilization filtration applications. The hydrophobic AerventPTFE membrane has high flow rate and flux, low precipitation and particle shedding, and wide chemical compatibility.

For easy traceability and identification, each Aervent-50 filter is engraved with a product number and batch number, and each Aervent-50 has passed integrity testing during production.

Usage: Used as a sterile respirator for containers and large glass bottles, a sterile respirator for high-pressure sterilization in a vacuum environment, a sterilization filter for non-aqueous solutions, a sterile respirator for small fermentation tanks, rolling bottles, cell factories, and low flow gas filtration.

Shell material:polypropylene

Filter membrane material:PTFE (polytetrafluoroethylene)

Membrane area:19.6 cm2

Maximum intake pressure:4.1 bar (60 psig) /25C

Bacterial interception:Using Brevundimonas diminuta (ATCC 19146) following ASTM 838-83. technique, challenge the bacterial retention rate of the filter at a minimum bacterial level of 107 CFU/cm2

Bacterial endotoxins:LAL detection value ≤ 05EU/mL

Weight analysis of detachment and precipitation:≤ 10mg/piece, using 70/30% isopropanol (IPA)/water

Sterilization:10 times 130 ° C 30 minute high-pressure sterilization

Integrity testing bubble point:≥1 100 mbar (16 psig), Use 60/40% isopropanol (IPA)/water/23 ° C