Online hardness analyzer

NegotiableUpdate on 04/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The GL-7200 online hardness analyzer is compact in structure and easy to operate. It is a device developed based on the principle of national standard hardness titration method to automatically monitor residual hardness in water quality. Equipped with Greencaire's self-developed color grading mode algorithm, it can automatically eliminate the interference of external light changes, ignore the turbidity and color interference of the original water sample, and the pollution interference caused by long-term measurement of the colorimetric pool. It can automatically and accurately determine the hardness endpoint, ensuring the reliability of real-time data.
Product Details

Product Overview


GL-7200Online hardness analyzerThe device is compact in structure and easy to operate, developed based on the principle of national standard hardness titration method, and automatically monitors residual hardness in water quality.

Equipped with Greenery's self-developed color grading mode algorithm, it can automatically eliminate external light interference, ignore the turbidity and color interference of the original water sample, and the colorimetric effect caused by long-term measurement The interference of pool pollution automatically and accurately determines the hardness endpoint, ensuring the reliability of real-time data.GL-7200Online hardness analyzerThe equipment can set alarm and control alarm values, providing important data support for process control. It is an ideal choice for water quality monitoring in power plants, pharmaceutical factories, purified water systems, and industrial boiler water.


Reagent model and measurement range


model

mg/L(以CacO3 counts)

Ppm (calculated as Cac03)

mmol/L

GLSOLTH-01

0.2-2

0.2-2

0.002-0.02

GLS-OLTH-02

0.5-5

0.5-5

0.005-0.05

GLS-OLTH-03

1-10

1-10

0.01-0.1

GLS-OLTH-04

5-50

5-50

0.05-0.5

GLS-OLTH-05

10-100

10-100

0.1-1

GLS-OLTH-06

25-250

25-250

0.25-2.5

GLS-OLTH-07

50-500

50-500

0.5-5

Features


Automatically measure the total hardness of water within the corresponding reagent kit range. More efficient than manual measurement and more stable and reliable compared to ion selective electrodes.

Based on the national standard titration methodThe development of hardness determination in GB/T6909-2018 has changed from red to blue as the endpoint.

Built in calibration function, users can calibrate themselves.

It can achieve continuous measurement, interval measurement, and external signal triggered measurement.

3.5-inch color screen display, supporting both Chinese and English languages.

The interface adopts a guided menu design, which can quickly get started and use the device.

A 450mL reagent kit can support 2000-4500 measurements.

It can automatically eliminate external light interference, water sample turbidity and color interference, and colorimetric pool pollution interference caused by long-term analysis, and automatically and accurately determine the hardness endpoint.

After each analysis is completed, automatic flushing will be performed, and the flushing duration can be customized to ensure measurement accuracy and repeatability, reducing on-site maintenance.

2 sets of programmable relay outputs, 1 set of 0/4-20mA outputs.

size300x300x170mm, Easy to hang and install.

About storageOne million pieces of data can be exported to a computer to view historical test data.

Can set and control alarm values or trigger external signals when the alarm value is reached


technical indicators


measurement method

Titration colorimetric method

measurement range

0~500mg/L (segmented according to the type of reagent)

measure time

about3-5 minutes per session

indication error

Selected reagents±5%

repetitiveness

Selected reagents5%

reagent consumption

about0.2mL/time, depending on concentration

Validity period of reagents

2 years, stored at room temperature and away from light.

Water sample consumption

about1L water sample, depending on water sample pressure and flushing time

power supply

100-240VAC,50/60Hz

power consumption

25W

Protection level

IP65

display

3.5-inch color screen

selectable unit

mg/LCaCO3, ppmCaC03, mmol/L, mval/L

input

2 sets of programmable inputs IN1, IN2 (start analysis/stop analysis/instrument reset)

output

2 sets of programmable relay outputs (max. 250V. 4A)

1 set of 0/4~20mA signals, max.7502

RS485 Modbus RTU communication

Analysis cycle

10 minutes to 720 minutes

Rinse time

10 seconds to 1800 seconds

water quality requirements

Colorless, no suspended solids, no bubbles

General surface water, boiler water, tap water

temperature

5~40℃

humidity

≤ 85% non condensing

pressure

0.5-6 bar (recommended 1-2 bar)

enter/Water outlet connection

outer diameter6mm hose

size

300x300x170mm

weight

4kg

Installation method

wall-mounted