Radcal Dynalyzer IV dynamic analyzer detector from the United StatesRepresented by iInstrument Network, Dynalyzer IV is one of the most accurate tools for calibrating and analyzing X-ray generator systems. Currently on sale, if you need to make a purchase, you can contact us through the customer service hotline of ai1718!

Radcal Dynalyzer IV dynamic analyzer detector from the United StatesProduct Features:
DYNALYZER IV High Voltage Unit
Dynalyzer IV is one of the most accurate tools for calibrating and analyzing X-ray generator systems. The X-ray generator of Dynalyzer IV is connected to a high-voltage cable (located between the X-ray generator and the X-ray tube), which provides analog and digital signals for grounding reference. These signals are proportional to kilovolts (kV), anode current (mA), and filament current. The analog versions of these signals can be routed to external testing equipment (such as oscilloscopes) or converted into digital voltage signals. The latest version of the device (anode cathode voltage and anode current versions) can be connected to the Dynalyzer Accu Dyn+display system and used in conjunction with Accu Gold Excel software. Unlike non-invasive instruments, Dynalyzer is not affected by tube target materials or filtering devices and can provide accurate kilovolt measurement results.

DYNALYZER ACCU-DYN+display system
The Accu Dyn+display system provides a way to display, analyze, and report Dynalyzer measurement data, which can support calibration, compliance testing, and troubleshooting. The displayed calibration data includes: kilovolt value, peak tube voltage (kVp), anode current, number of pulses, and pulse duration. In addition, it can also display waveforms of kilovolts and anode current, allowing users to have a deeper understanding of the system's operating status.
Dynalyzer paired with Accu Dyn+display system:
| core functionality | Corresponding advantages |
| auto-trigger | After enabling the "automatic triggering" mode, real-time measurement of generator parameters can be carried out |
| Custom Report | Accu Gold Excel supports users to create automated reports that meet their own needs |
| print report | The Windows display can connect to any Windows compatible printer for report printing |
| Advanced Computing | Automatically calculate peak voltage (PPV), peak tube voltage (kVp), and average kilovolt value (kV) |
| Report Archiving | The report is stored in Excel format and can be accessed at any time as needed |
Accessories and Accessories Selection
(For other optional accessories and attachments, please refer to the price list)
Dynalyzer IV High Voltage Unit Special Bracket
Suitable for 3-conductor cables and 12 foot coaxial cables, including 2 plugs; The bracket for 4-conductor cables needs to be specially ordered
Heavy transport box (M-96115)
Compatible with Dynalyzer IV (requires separate order)
High voltage cable (M-53605)
Suggest equipping 2 5-foot long cables
Measurable parameters:
Peak tube voltage (kVp): including three measurement dimensions: anode, cathode, and anode+cathode
Anode current: milliampere (mA) and milliampere second (mAs)
Filament current
exposure time
Applicable scenarios:
Suitable for the production, manufacturing, troubleshooting, annual calibration, installation, commissioning, and acceptance testing of X-ray machines
Technical parameters of Radcal Dynalyzer IV dynamic analyzer from the United States:
| 1、 Dynalyzer IV High Voltage Unit Technical Parameters | ||
| parameter item | precision | measurement range |
| Kilovolt voltage accuracy | ±1% | 10 kV to 150 kV |
| Anode current | ±1% | 1 mA to 1 A (DC to 4 kHz) |
| Filament current | ±5% | DC to 20 kHz |
| Note: All specifications may be subject to change. | ||
| 2、 Accu Dyn+Display System Technical Parameters | ||
| parameter item | precision | measurement range |
| Kilovolt voltage accuracy | ±1% | 10-150 kV (DC to 4.5 kHz) |
| Anode current | ±0.2% | 0.1 mA to 1 A (DC to 23 kHz) |
| exposure time | ±0.5 msec | - |
| Note: The waveform and calculated kilovolt values reflect a bandwidth of 4.5 kHz. |