Simple operation and testing of dental module brush grinder
1、 Instrument Introduction:
Suitable for testing the cleaning rate of external color spots in toothpaste. Through experiments, it is possible to effectively compare and evaluate the relative cleaning rate of toothpaste for exogenous pigmentation, providing scientific basis for determining relevant technical indicators.
2、 Key parameters
Parameter items |
Specifications |
control system |
Industrial grade PLC |
operating interface |
Colorful 7-inch touch screen, switching between Chinese and English |
Operation method |
Full touch operation, capable of setting, calling, and modifying test programs |
printer |
Embedded printer |
Testing station |
9 workstations |
Test count |
0 times -99999 times |
test frequency |
0r/min-150r/min (adjustable) |
Testing direction |
X. The Y-axis direction can be changed |
Vertical pressure |
150 ± 2g can be optionally equipped with other pressures |
security configuration |
Emergency stop button, overload protection |
power supply |
AC 220V/50Hz |
Overall dimensions |
Approximately 680 * 750 * 1100mm |
USB testing software socket, capable of reading data and connecting to a computer | |
3、 Execution standards and applicable scenarios
meet the standards
GB/T43576-2023 Laboratory Test Method for Toothpaste Effectiveness in Removing Exogenous Stains in Oral Cleaning and Care Products
Batch quality inspection;
Third party laboratory certification.
4、 Example of Operation Process
Install the standard toothbrush and dental grinding block onto the grinding device, adjust the vertical pressure of the toothbrush on the dental block to 150g ± 2g, and then set the back and forth frequency of the grinding device to 100r/min. Add toothpaste slurry and control abrasive slurry to the slurry cups of the sample group and control group, respectively. Add an equal volume of slurry to each slurry cup to ensure that the slurry completely submerges the dental grinding block. Start the grinding device and make 600 round trips to the teeth.
After brushing and grinding, remove the dental grinding block and rinse it with water three times until there is no visible tooth pulp residue. Dry the surface moisture with paper and measure the enamel whiteness value (L) within 10 minutes using a colorimeter.
Note: A new toothbrush needs to be replaced during each experiment. There may be sedimentation phenomenon in the few plate samples, and the dental pulp should be stirred during the process.