4010 gas dilution calibrator

NegotiableUpdate on 01/05
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Instrument Introduction:
Product Details

Instrument Introduction:

Users can conveniently calibrate gas dilution, ozone, and GPT using the 4010 gas dilution calibrator. The calibration system of the 4010 instrument is convenient and effective in controlling the dilution of standard gases, and is used for the calibration of environmental air and pollution source analyzers.

The 4010 calibrator uses built-in microprocessor technology to enhance the accuracy of the instrument and control over the calibration system. The instrument can be manually, automatically, or remotely controlled. The calibrator can generate and store 20 calibration sequences, each with up to 20 dilution ratios. Users can input calibration parameters using the buttons on the front panel and external keyboard, and control the calibration action through a series of interfaces.

The gas path system of the 4010 calibrator uses a mass flow controller to measure gas flow rate for linear calibration.

The design of the 4010 calibrator exceeds the requirements of the EPA standard calibration method. All mass flow controllers are calibrated through NIST's standardized transmission.

Standard configuration:

Gas path section

*Dilution gas mass flow controller, 0-10 SLM;

*Source gas mass flow controller, 0-100 SCCM;

*4 calibration gas input interfaces and 1 cleaning interface;

*5 output interfaces and 1 air outlet;

*1 dilution gas input interface;

controller

*High brightness color display screen (displayed as black and white EL screen);

*55 key touch keyboard;

*24 bit digital I/O interface;

*2 serial data communication interfaces;

*External PC keyboard input interface;

*Parallel printer interface;

*Air source control interface;

Power supply section:

*Standard 98-264 VAC, 150-300 VA, 50/60 Hz operation;

Optional configuration:

*Dilution gas mass flow controller, range 0-20 SLM;

*Source gas mass flow controller, with a range of 0-1 SLM;

*The second source gas mass flow controller has a range of 0-2 SLM;

*Optional additional calibration gas inlet;

*External infiltration furnace equipment interface;

*Optional additional output interfaces;

*The second dilution gas inlet;

*Built in ozone generator;

*Ozone generator UV closed-loop control system;

*Built in ozone UV absorption spectrophotometer;

*Output solenoid;

*Provide 6 external solenoid valve drive ports, drive power: 24VDC/3w;

*Installation bracket;

*Ethernet TCP/IP interface;

Product specifications:

Dilution system

*Dilution gas inlet: 1 standard interface (2 optional interfaces);

*Source gas inlet: 4 standard interfaces, 1 cleaning interface;

*Optional import: can add 2 interfaces;

*Dilution gas mass flow controller: 0-10 SLPM; Optional range 0-20 SLPM;

*Source gas mass flow controller: 0-100 SCCM; Optional range 0-1 SLPM;

*Second source gas mass flow controller: 0-2 SLPM;

*Flow accuracy: ≤± 1% of full scale;

*Flow reproducibility: ≤± 0.15% of full scale;

*Linear: ≤± 0.5% of full scale;

*Input pressure: 20-35 PSI;

*Output interface: 6 standard output ports;

*Optional input interfaces: 2;

Optional built-in ozone generator:

*Output: 0.05-1.5 ppm@5 SLPM;

*Accuracy: ± 2% or ± 3 ppb@5 SLPM;

*Flow rate: 100 SCCM, ± 1 SCCM;

*UV lamp temperature: 50 ℃, ± 0.1 ℃;

Optional built-in infiltration furnace:

*Box material: glass or Teflon;

*Box size: 1 heating tube;

*Adjustable temperature: ambient temperature above 5 ℃ to 65 ℃;

*Temperature accuracy: ≤± 0.1 ℃;

Calibration interface:

*Operation: Manually controlled through the 55 key touch keyboard;

*Calibration settings: 20 factory set calibration sequences, with a maximum of 20 calibration points per sequence;

*Calibration type: gas dilution, ozone, GPT;

*Gas setting: 2 dilution gases, 20 source gases;

*Automatic calibration: Provide 20 timed calibration paths based on user set calibration events within 7 days;

*Digital access port: 24 bit control calibration function;

*Digital interface type: closed-loop connection or TTL logical connection;

*Digital output port: 24 bit output monitoring and calibration function;

*Interface: RS232;

Calibration system:
*Operating (environmental) temperature: 5-40 ℃;
*Specification size: 8.75 "(22.2 cm) H × 17" (43.2 cm) W × 20 "(50.8 cm) D;
*Weight: 40 lbs. (18.1 kg);
*Input voltage: 98-264 VAC, 150-300 VA, operating at 50/60 Hz.