The extraction type ultra-low dust instrument online monitoring system has been widely used to reduce the concentration of fine suspended particles through wet electrostatic precipitator technology. To measure the concentration of high humidity dust after wet electrostatic precipitator, an extraction type dust analyzer is selected. The $n extraction dust meter is mainly used for measuring the concentration of dust with high humidity and viscosity. The instrument extracts flue gas with high moisture content from the flue, removes moisture through heating, and measures dust concentration. In addition to heating the flue gas, the $n extraction dust meter also adds a heated dilution gas to heat and dilute the original flue gas, ensuring that only dust particles enter the measurement
Product Details

Extraction based ultra-low dust instrument online monitoring system

The ultra-low smoke detector is based on the principle of forward light scattering. The highly stable laser emitted by the laser is collimated and enters the measurement chamber. When the laser beam meets the particles, forward scattered light is generated. The scattered light is received by the lens and introduced into the photodetector through the signal fiber. After the ultra-high sensitivity photoelectric signal conversion, the scattered light signal is transformed into an electrical signal proportional to the concentration of the particles. Through a specific algorithm, the concentration value of the particles is output.

Working principle: Laser forward scattering
Measurement objects: smoke and particulate matter
Measurement range: small (0-5) (mg/m3), (0-200) (mg/m3)
Zero and range drift: ≤± 1.0% F.S./week
Resolution: 0.005mg/m3
Analog output: 4-20 (mA) (output load 500)
Digital output: RS485 (supporting Modbus protocol)
Flue diameter: (0.3-20) meters
Smoke flow velocity: (0-30) m/s
Backblowing setting: Automatic (Backblowing time interval can be set by oneself)
Sampling method: extraction based
Tested medium conditions: 10 ℃~300 ℃ (customized for high temperatures)
Protection level: IP65 for the system
Power supply: 220 ± 10% (VAC), 500W
Overall dimensions:
Host: 410mm * 502mm * 230mm
Constant speed sampling box: 404mm x 300mm x 140mm
Weight: Host ≤ 15kg, constant velocity sampling box ≤ 7kg

Extraction based ultra-low dust instrument online monitoring system

Has high sensitivity and resolution; Extraction heating return gas path structure, which can be used in situations where the measured gas contains high water mist; Attached is an isokinetic tracking sampling module, which can be configured with isokinetic sampling; Adopting an adaptive dynamic grading method with a wide dynamic response range; *Automatic zero point and automatic span calibration; The structure and calculation method of the blowback and air curtain held, resistant to window pollution; *The structural design includes an integrated host and a separate isokinetic tracking module. The overall weight of the host is less than 20KG, and a single person can complete installation, debugging, and maintenance work.

Range: 0~10mg/m3 (Extractive ultra-low emission dust meter) Linearity: ± 2% FS

Zero drift:<± 2% FS/week
Range drift:<± 2% FS/week

Accuracy:<± 2% FS Response time: 5S Installation method: bypass extraction type

Monitoring principle: Laser scattering method. Low detection limit: 0.1mg/m3. Environmental temperature: -20~60 ° C. Using air: 0.4~0.6Mpa. Output signal: 4~20mA